Fluorescence and reflectance device variability throughout the progression of a phase II clinical trial to detect and screen for cervical neoplasia using a fiber optic probe
- PMID: 17614723
- DOI: 10.1117/1.2750332
Fluorescence and reflectance device variability throughout the progression of a phase II clinical trial to detect and screen for cervical neoplasia using a fiber optic probe
Abstract
Large phase II trials of fluorescence and reflectance spectroscopy using a fiber optic probe in the screening and diagnostic settings for detecting cervical neoplasia have been conducted. We present accrual and histopathology data, instrumentation, data processing, and the preliminary results of interdevice consistencies throughout the progression of a trial. Patients were recruited for either a screening trial (no history of abnormal Papanicolaou smears) or a diagnostic trial (a history of abnormal Papanicolaou smears). Colposcopy identified normal and abnormal squamous, columnar, and transformation zone areas that were subsequently measured with the fiber probe and biopsied. In the course of the clinical trial, two generations of spectrometers (FastEEM2 and FastEEM3) were designed and utilized as optical instrumentation for in vivo spectroscopic fluorescence and reflectance measurements. Data processing of fluorescence and reflectance data is explained in detail and a preliminary analysis of the variability across each device and probe combination is explored. One thousand patients were recruited in the screening trial and 850 patients were recruited in the diagnostic trial. Three clinical sites attracted a diverse range of patients of different ages, ethnicities, and menopausal status. The fully processed results clearly show that consistencies exist across all device and probe combinations throughout the diagnostic trial. Based on the stratification of the data, the results also show identifiable differences in mean intensity between normal and high-grade tissue diagnosis, pre- and postmenopausal status, and squamous and columnar tissue type. The mean intensity values of stratified data show consistent separation across each of the device and probe combinations. By analyzing trial spectra, we provide more evidence that biographical variables such as menopausal status as well as tissue type and diagnosis significantly affect the data. Understanding these effects will lead to better modeling parameters when analyzing the performance of fluorescence and reflectance spectroscopy.
Similar articles
-
Instrumentation as a source of variability in the application of fluorescence spectroscopic devices for detecting cervical neoplasia.J Biomed Opt. 2007 May-Jun;12(3):034014. doi: 10.1117/1.2745285. J Biomed Opt. 2007. PMID: 17614722
-
Effect of probe pressure on cervical fluorescence spectroscopy measurements.J Biomed Opt. 2004 May-Jun;9(3):523-33. doi: 10.1117/1.1695562. J Biomed Opt. 2004. PMID: 15189090 Clinical Trial.
-
The performance of fluorescence and reflectance spectroscopy for the in vivo diagnosis of cervical neoplasia; point probe versus multispectral approaches.Gynecol Oncol. 2007 Oct;107(1 Suppl 1):S248-55. doi: 10.1016/j.ygyno.2007.07.008. Epub 2007 Sep 7. Gynecol Oncol. 2007. PMID: 17825399 Review.
-
Effects of biographical variables on cervical fluorescence emission spectra.J Biomed Opt. 2003 Jul;8(3):479-83. doi: 10.1117/1.1578642. J Biomed Opt. 2003. PMID: 12880354
-
The clinical effectiveness of fluorescence and reflectance spectroscopy for the in vivo diagnosis of cervical neoplasia: an analysis by phase of trial design.Gynecol Oncol. 2007 Oct;107(1 Suppl 1):S270-80. doi: 10.1016/j.ygyno.2007.07.009. Epub 2007 Sep 7. Gynecol Oncol. 2007. PMID: 17825882 Review.
Cited by
-
Early detection of high-grade squamous intraepithelial lesions in the cervix with quantitative spectroscopic imaging.J Biomed Opt. 2013 Jul;18(7):76013. doi: 10.1117/1.JBO.18.7.076013. J Biomed Opt. 2013. PMID: 23843090 Free PMC article.
-
Prospective detection of cervical dysplasia with scanning angle-resolved low coherence interferometry.Biomed Opt Express. 2020 Aug 20;11(9):5197-5211. doi: 10.1364/BOE.401000. eCollection 2020 Sep 1. Biomed Opt Express. 2020. PMID: 33014608 Free PMC article.
-
Accuracy of optical spectroscopy for the detection of cervical intraepithelial neoplasia without colposcopic tissue information; a step toward automation for low resource settings.J Biomed Opt. 2012 Apr;17(4):047002. doi: 10.1117/1.JBO.17.4.047002. J Biomed Opt. 2012. PMID: 22559693 Free PMC article.
-
Physician attitudes toward dissemination of optical spectroscopy devices for cervical cancer control: an industrial-academic collaborative study.Gend Med. 2012 Feb;9(1 Suppl):S67-77; quiz 77.e1-6. doi: 10.1016/j.genm.2011.11.004. Gend Med. 2012. PMID: 22340642 Free PMC article.
-
Effect of anatomy on spectroscopic detection of cervical dysplasia.J Biomed Opt. 2009 Jul-Aug;14(4):044021. doi: 10.1117/1.3194142. J Biomed Opt. 2009. PMID: 19725732 Free PMC article.