A new endoluminal, flow-disrupting device for treatment of saccular aneurysms
- PMID: 17615366
- DOI: 10.1161/STROKEAHA.106.479576
A new endoluminal, flow-disrupting device for treatment of saccular aneurysms
Abstract
Background and purpose: We report a preclinical study of a new endoluminal device for aneurysm occlusion to test the hypothesis that the device, even without use of intrasaccular coil placement, could occlude saccular aneurysms without causing substantial parent artery compromise or compromise of adjacent, small branch arteries.
Methods: The Pipeline Neuroendovascular Device (Pipeline NED; Chestnut Medical Technologies, Inc) is a braided, tubular, bimetallic endoluminal implant aimed at occlusion of saccular aneurysms through flow disruption along the aneurysm neck. The device was implanted across the necks of 17 elastase-induced aneurysms in the New Zealand white rabbit model and followed for 1 month (n=6), 3 months (n=5), and 6 months (n=6). In each subject, a second device was implanted in the abdominal aorta to cover the origins of lumbar arteries. Aneurysm occlusion rates by angiography (grade 1, complete occlusion; grade 2, near-complete occlusion; and grade 3, incomplete occlusion) were documented. Percent area stenosis of the parent arteries was calculated. Presence of distal emboli in the downstream vessels in the parent artery and branch artery stenosis or occlusion was noted.
Results: Grades 1, 2, and 3 occlusion rates were noted in 9 (53%), 6 (35%), and 2 (12%) of 17 aneurysms, respectively, indicating an 88% rate of complete or near complete occlusion. No cases of branch artery occlusion or distal emboli in the downstream vessels of the parent artery, specifically the subclavian artery, were seen. Parent artery compromise from neointimal hyperplasia was minimal in most cases.
Conclusions: The Pipeline NED is a trackable, bio- and hemocompatible device able to occlude saccular aneurysms with preservation of the parent artery and small, adjacent branch vessels.
Similar articles
-
A second-generation, endoluminal, flow-disrupting device for treatment of saccular aneurysms.AJNR Am J Neuroradiol. 2009 Jun;30(6):1153-8. doi: 10.3174/ajnr.A1530. Epub 2009 Apr 15. AJNR Am J Neuroradiol. 2009. PMID: 19369609 Free PMC article.
-
The Derivo Embolization Device, a Second-Generation Flow Diverter for the Treatment of Intracranial Aneurysms, Evaluated in an Elastase-Induced Aneurysm Model.Clin Neuroradiol. 2017 Sep;27(3):335-343. doi: 10.1007/s00062-015-0493-9. Epub 2015 Dec 23. Clin Neuroradiol. 2017. PMID: 26699192
-
Treatment of distal anterior cerebral artery aneurysms with the Pipeline Embolization Device.J Clin Neurosci. 2017 Jan;35:133-138. doi: 10.1016/j.jocn.2016.10.041. Epub 2016 Nov 15. J Clin Neurosci. 2017. PMID: 27863970
-
Curative cerebrovascular reconstruction with the Pipeline embolization device: the emergence of definitive endovascular therapy for intracranial aneurysms.J Neurointerv Surg. 2018 Jul;10(Suppl 1):i9-i18. doi: 10.1136/jnis.2009.000083.rep. J Neurointerv Surg. 2018. PMID: 30037946 Review.
-
Endovascular management of intracranial aneurysms.Neurosurg Clin N Am. 2009 Oct;20(4):383-98. doi: 10.1016/j.nec.2009.07.003. Neurosurg Clin N Am. 2009. PMID: 19853799 Review.
Cited by
-
Evaluation of occurring complications after flow diverter treatment of elastase-induced aneurysm in rabbits using micro-CT and MRI at 9.4 T.Neuroradiology. 2016 Oct;58(10):987-996. doi: 10.1007/s00234-016-1730-x. Epub 2016 Jul 25. Neuroradiology. 2016. PMID: 27457491
-
Adaptive Remodeling in the Elastase-induced Rabbit Aneurysms.Exp Mech. 2021 Jan;61(1):263-283. doi: 10.1007/s11340-020-00671-9. Epub 2020 Oct 27. Exp Mech. 2021. PMID: 33814553 Free PMC article.
-
Treatment of a pediatric recurrent fusiform middle cerebral artery (MCA) aneurysm with a flow diverter.BMJ Case Rep. 2012 Nov 15;2012:bcr2012010478. doi: 10.1136/bcr-2012-010478. BMJ Case Rep. 2012. PMID: 23162037 Free PMC article.
-
Risk Factors for Ischemic Complications following Pipeline Embolization Device Treatment of Intracranial Aneurysms: Results from the IntrePED Study.AJNR Am J Neuroradiol. 2016 Sep;37(9):1673-8. doi: 10.3174/ajnr.A4807. Epub 2016 Apr 21. AJNR Am J Neuroradiol. 2016. PMID: 27102308 Free PMC article.
-
An in vitro study of silk stent morphology.Neuroradiology. 2011 Sep;53(9):659-67. doi: 10.1007/s00234-010-0784-4. Epub 2010 Oct 28. Neuroradiology. 2011. PMID: 20981414
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical