Reconstructed versus real-time 3-dimensional echocardiography: comparison with magnetic resonance imaging
- PMID: 17617313
- DOI: 10.1016/j.echo.2006.12.010
Reconstructed versus real-time 3-dimensional echocardiography: comparison with magnetic resonance imaging
Abstract
Background: The accuracy, reproducibility, and test-retest reliability of 3-dimensional (3D) echocardiography (3DE) with 3D reconstruction (3DR) and real-time (RT) imaging (RT 3DE) exceed that of 2-dimensional echocardiography (2DE). However, image quality with RT 3DE is inferior to 2DE and we sought to determine whether this justified ongoing use of 3DR.
Methods: Unselected patients (n = 30, 22 men, age 66 +/- 7 years) presenting to the echocardiography laboratory for left ventricular (LV) evaluation were studied with 2DE and RT 3DE; 3DR images were obtained using external localization. The 3D measurements and reconstructions were obtained offline. Magnetic resonance images (MRI) were obtained using true free induction, steady state precession during breath hold and 3D volumes and ejection fraction (EF) were measured using 3D software. A separate cohort of 20 patients (13 men, age 60 +/- 12 years) was measured for test-retest variation.
Results: All echocardiographic measures underestimated LV volumes and EF compared with MRI, but this was least with RT 3DE. End-diastolic volume by MRI (168 +/- 54 mL) was underestimated by RT 3DE (-15 +/- 31, P = .02), 3DR (-26 +/- 33, P < .01), and 2DE (-57 +/- 40, P < .01). Similarly, end-systolic volume by MRI (86 +/- 50 mL) was underestimated by RT 3DE (-15 +/- 31, P = .02), 3DR (-26 +/- 33, P < .01), and 2DE (-57 +/- 40, P < .01). However, EF measurements were similar with each method. Test-retest variation was less and interobserver and intraobserver correlations were better with RT 3DE for volumes and EF, compared with 3DR and 2DE.
Conclusions: Despite limitations of image quality, RT 3DE is the most feasible and accurate approach for LV volume and EF measurements and follow-up LV assessment in daily practice.
Similar articles
-
Quantification of left ventricular volumes and ejection fraction using freehand transthoracic three-dimensional echocardiography: comparison with magnetic resonance imaging.J Am Soc Echocardiogr. 2003 Feb;16(2):101-9. doi: 10.1067/mje.2003.7. J Am Soc Echocardiogr. 2003. PMID: 12574735
-
Rapid full volume data acquisition by real-time 3-dimensional echocardiography for assessment of left ventricular indexes in children: a validation study compared with magnetic resonance imaging.J Am Soc Echocardiogr. 2005 Apr;18(4):299-305. doi: 10.1016/j.echo.2004.11.002. J Am Soc Echocardiogr. 2005. PMID: 15846155
-
What is the optimal clinical technique for measurement of left ventricular volume after myocardial infarction? A comparative study of 3-dimensional echocardiography, single photon emission computed tomography, and cardiac magnetic resonance imaging.J Am Soc Echocardiogr. 2006 Feb;19(2):192-201. doi: 10.1016/j.echo.2005.08.020. J Am Soc Echocardiogr. 2006. PMID: 16455424 Clinical Trial.
-
Three-dimensional echocardiography in evaluation of left ventricular indices.Echocardiography. 2012;29(1):66-75. doi: 10.1111/j.1540-8175.2011.01526.x. Echocardiography. 2012. PMID: 23186291 Review.
-
Imaging the heart failure patient-need for accurate measurements of left ventricular volumes and ejection fraction: the role of three-dimensional and contrast echocardiography.Curr Opin Cardiol. 2016 Sep;31(5):459-68. doi: 10.1097/HCO.0000000000000312. Curr Opin Cardiol. 2016. PMID: 27271926 Review.
Cited by
-
Feasibility of real-time three-dimensional echocardiography for the assessment of distorted biventricular systolic function in patients with cor pulmonale.J Cardiovasc Ultrasound. 2013 Jun;21(2):64-71. doi: 10.4250/jcu.2013.21.2.64. Epub 2013 Jun 26. J Cardiovasc Ultrasound. 2013. PMID: 23837116 Free PMC article.
-
Diastolic pressure-volume quotient (DPVQ) as a novel echocardiographic index for estimation of LV stiffness in HFpEF.Clin Res Cardiol. 2015 Nov;104(11):955-63. doi: 10.1007/s00392-015-0863-y. Epub 2015 May 9. Clin Res Cardiol. 2015. PMID: 25956143
-
[Real-time 3 dimensional full volume data set : benefits in problem focused intraoperative transesophageal echocardiography].Anaesthesist. 2012 Oct;61(10):875-82. doi: 10.1007/s00101-012-2088-z. Epub 2012 Oct 20. Anaesthesist. 2012. PMID: 23080355 German.
-
Non-Invasive Evaluation of Heart Function with Four-Dimensional Echocardiography.PLoS One. 2016 May 4;11(5):e0154996. doi: 10.1371/journal.pone.0154996. eCollection 2016. PLoS One. 2016. PMID: 27144844 Free PMC article.
-
Three Dimensional (3D) Echocardiography as a Tool of Left Ventricular Assessment in Children with Dilated Cardiomyopathy: Comparison to Cardiac MRI.Open Access Maced J Med Sci. 2018 Oct 25;6(12):2310-2315. doi: 10.3889/oamjms.2018.270. eCollection 2018 Dec 20. Open Access Maced J Med Sci. 2018. PMID: 30607182 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical