Evaluation of a limited cone-beam volumetric imaging system: comparison with film radiography in detecting incipient proximal caries
- PMID: 17618136
- DOI: 10.1016/j.tripleo.2007.02.028
Evaluation of a limited cone-beam volumetric imaging system: comparison with film radiography in detecting incipient proximal caries
Abstract
Objectives: To assess the accuracy of a limited cone-beam volumetric imaging system in detecting incipient proximal caries.
Study design: Fifty extracted premolars were used. On 100 surfaces examined, 71 proximal carious lesions were found; 47 were limited to the enamel, and 24 penetrated into the dentin. A limited cone-beam volumetric imaging system (CBVI), 3D Accuitomo (J. Morita Mfg. Corp., Kyoto, Japan), was used. Insight films (Eastman Kodak, Rochester, NY) were used for comparison. Seven experienced oral radiologists determined the presence of lesions using a continuous rating scale. The area measure (Az) values were calculated from the receiver operating characteristic curve obtained for each modality and observer.
Results: The mean Az-values were 0.63 +/- 0.02 for 3D Accuitomo and 0.63 +/- 0.03 for film. Analysis of variance showed no statistically significant difference between 3D Accuitomo and film.
Conclusions: A limited CBVI, 3D Accuitomo, could not enhance the accuracy in detecting proximal carious lesions.
Similar articles
-
A comparative study of different radiographic methods for detecting occlusal caries lesions.Caries Res. 2014;48(6):566-74. doi: 10.1159/000357596. Epub 2014 Jul 29. Caries Res. 2014. PMID: 25073755
-
Detection accuracy of in vitro approximal caries by cone beam computed tomography images.Eur J Radiol. 2011 Aug;79(2):e24-7. doi: 10.1016/j.ejrad.2009.05.063. Epub 2010 Feb 24. Eur J Radiol. 2011. PMID: 20185260
-
An in-vitro evaluation of Kodak Insight and Ektaspeed Plus film with a CMOS detector for natural proximal caries: ROC analysis.Caries Res. 2001 Sep-Oct;35(5):354-9. doi: 10.1159/000047474. Caries Res. 2001. PMID: 11641571
-
Detection accuracy of proximal caries by phosphor plate and cone-beam computerized tomography images scanned with different resolutions.Clin Oral Investig. 2012 Aug;16(4):1015-21. doi: 10.1007/s00784-011-0599-7. Epub 2011 Jul 30. Clin Oral Investig. 2012. PMID: 21805053
-
Diagnostic accuracy of cone beam computed tomography compared with intraoral radiography for the detection of noncavitated occlusal carious lesions.Caries Res. 2014;48(5):461-6. doi: 10.1159/000358101. Epub 2014 May 21. Caries Res. 2014. PMID: 24852420
Cited by
-
The influence of amalgam fillings on the detection of approximal caries by cone beam CT: in vitro study.Dentomaxillofac Radiol. 2014;43(7):20130342. doi: 10.1259/dmfr.20130342. Epub 2014 Jul 2. Dentomaxillofac Radiol. 2014. PMID: 24986630 Free PMC article.
-
Radiographic modalities for diagnosis of caries in a historical perspective: from film to machine-intelligence supported systems.Dentomaxillofac Radiol. 2021 Jul 1;50(5):20210010. doi: 10.1259/dmfr.20210010. Epub 2021 Mar 4. Dentomaxillofac Radiol. 2021. PMID: 33661697 Free PMC article. Review.
-
Effect of Filtration and Thickness of Cross-Sections of Cone Beam Computed Tomography Images on Detection of Proximal Caries.J Dent (Tehran). 2017 Jul;14(4):223-230. J Dent (Tehran). 2017. PMID: 29285032 Free PMC article.
-
Use of Digital Diagnostic Aids for Initial Caries Detection: A Review.Dent J (Basel). 2023 Sep 28;11(10):232. doi: 10.3390/dj11100232. Dent J (Basel). 2023. PMID: 37886917 Free PMC article. Review.
-
Comparison of Two-Dimensional and Three-Dimensional Radiographs Using Clinically Relevant Parameters.Dent J (Basel). 2019 May 1;7(2):50. doi: 10.3390/dj7020050. Dent J (Basel). 2019. PMID: 31052379 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical