Effect of relative humidity on the electrostatic charge properties of dry powder inhaler aerosols
- PMID: 17619997
- DOI: 10.1007/s11095-007-9377-2
Effect of relative humidity on the electrostatic charge properties of dry powder inhaler aerosols
Abstract
Purpose: At present, there is no published data examining the effect of relative humidity on the electrostatic charges of dry powder inhaler aerosols. The charging behaviour of two commercial products, Pulmicort and Bricanyl Turbuhalers, were investigated using an electrical low pressure impactor (ELPI).
Methods: ELPI was successfully modified to disperse the aerosols at 60 l/min. Four doses from each new inhaler were sampled at 15, 40, 65, and 90% RH. Particles deposited on the impactor stages according to their aerodynamic diameters and their charges were measured simultaneously by the electrometers. The drug in each size fraction was quantified using HPLC.
Results: Both products generated bipolar charges. The charging behaviour of the two types of inhaler showed different humidity dependence although the mass output was not significantly affected. The absolute specific charge of budesonide fine particles from Pulmicort was the lowest at 40% RH but increased at lower and higher RHs. In contrast, the terbutaline sulfate fine particles from Bricanyl followed the expected trend of charge reduction with increasing RH.
Conclusions: The distinct trends of charging of aerosols from Pulmicort and Bricanyl Turbuhalers was explained by differences in hygroscopicity and other physicochemical factors between the two drugs.
Similar articles
-
Effect of Relative Humidity on Bipolar Electrostatic Charge Profiles of dry Powder Aerosols.Pharm Res. 2017 Aug;34(8):1707-1715. doi: 10.1007/s11095-017-2178-3. Epub 2017 May 24. Pharm Res. 2017. PMID: 28540500
-
Aerosol electrostatics. I: Properties of fine powders before and after aerosolization by dry powder inhalers.Pharm Res. 1997 Jun;14(6):698-705. doi: 10.1023/a:1012181818244. Pharm Res. 1997. PMID: 9210184
-
Measuring Bipolar Charge and Mass Distributions of Powder Aerosols by a Novel Tool (BOLAR).Mol Pharm. 2015 Sep 8;12(9):3433-40. doi: 10.1021/acs.molpharmaceut.5b00443. Epub 2015 Aug 14. Mol Pharm. 2015. PMID: 26222019
-
Dry powder aerosol delivery systems: current and future research directions.J Aerosol Med. 2006 Spring;19(1):21-7. doi: 10.1089/jam.2006.19.21. J Aerosol Med. 2006. PMID: 16551211 Review.
-
A review of factors affecting electrostatic charging of pharmaceuticals and adhesive mixtures for inhalation.Int J Pharm. 2016 Apr 30;503(1-2):262-76. doi: 10.1016/j.ijpharm.2016.01.076. Epub 2016 Feb 2. Int J Pharm. 2016. PMID: 26836710 Review.
Cited by
-
Design of a device for simultaneous particle size and electrostatic charge measurement of inhalation drugs.Pharm Res. 2008 Nov;25(11):2488-96. doi: 10.1007/s11095-008-9660-x. Epub 2008 Jul 1. Pharm Res. 2008. PMID: 18592352
-
Investigation of electrostatic behavior of a lactose carrier for dry powder inhalers.Pharm Res. 2008 Dec;25(12):2822-34. doi: 10.1007/s11095-008-9651-y. Epub 2008 Jun 26. Pharm Res. 2008. PMID: 18581205
-
Administration of dry powders during respiratory supports.Ann Transl Med. 2021 Apr;9(7):596. doi: 10.21037/atm-20-3946. Ann Transl Med. 2021. PMID: 33987294 Free PMC article. Review.
-
Comparison of in vitro deposition of pharmaceutical aerosols in an idealized child throat with in vivo deposition in the upper respiratory tract of children.Pharm Res. 2014 Jun;31(6):1525-35. doi: 10.1007/s11095-013-1258-2. Epub 2014 Jan 7. Pharm Res. 2014. PMID: 24395403
-
Effect of crystallinity on electrostatic charging in dry powder inhaler formulations.Pharm Res. 2014 Jul;31(7):1656-64. doi: 10.1007/s11095-013-1270-6. Epub 2014 Jan 24. Pharm Res. 2014. PMID: 24464269
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources