Induction and prevention of pin loosening in external fixation: an in vivo study on sheep tibiae
- PMID: 1762013
- DOI: 10.1097/00005131-199112000-00017
Induction and prevention of pin loosening in external fixation: an in vivo study on sheep tibiae
Abstract
In external fixation of fractures, pin loosening is a major concern. Preloading the pins is generally done to ensure their stability within the bone cortex. The effect of radial preload and bending preload in reducing resorption at the pin/bone interface was tested. Schanz screws were fixed to live sheep tibiae using a pneumatically operated external fixator frame. Evaluation was based on radiological observation and fluorochrome histology using sequential labels. Though not completely absent, bone resorption was minimal in the radial preload group, compared with the two other groups. More important, the bone-to-pin contact surface in the radial preload group was found to be almost intact after 5 weeks. In this study, radial preload appears to be superior to bending preload in terms of minimizing the problems of pin loosening.
Similar articles
-
Axial preload in external fixator half-pins: a preliminary mechanical study of an experimental bone anchorage system.Clin Biomech (Bristol). 1999 Jan;14(1):69-73. doi: 10.1016/s0268-0033(98)00047-3. Clin Biomech (Bristol). 1999. PMID: 10619092
-
Analysis of the external fixator pin-bone interface.Clin Orthop Relat Res. 1993 Aug;(293):18-27. Clin Orthop Relat Res. 1993. PMID: 8339479
-
The effect of radial preload on the implant-bone interface: a cadaveric study.J Orthop Trauma. 1989;3(4):323-32. doi: 10.1097/00005131-198912000-00012. J Orthop Trauma. 1989. PMID: 2600701
-
Application and postoperative management of external skeletal fixators.Vet Clin North Am Small Anim Pract. 1992 Jan;22(1):69-97. doi: 10.1016/s0195-5616(92)50005-5. Vet Clin North Am Small Anim Pract. 1992. PMID: 1539431 Review.
-
Hydroxyapatite-coated external fixation pins.Expert Rev Med Devices. 2005 Jul;2(4):465-71. doi: 10.1586/17434440.2.4.465. Expert Rev Med Devices. 2005. PMID: 16293085 Review.
Cited by
-
Effects of Ti2448 half-pin with low elastic modulus on pin loosening in unilateral external fixation.J Mater Sci Mater Med. 2011 Jun;22(6):1579-88. doi: 10.1007/s10856-011-4313-8. Epub 2011 Apr 13. J Mater Sci Mater Med. 2011. PMID: 21487787
-
Computational simulations of stress shielding and bone resorption around existing and computer-designed orthopaedic screws.Med Biol Eng Comput. 2002 May;40(3):311-22. doi: 10.1007/BF02344213. Med Biol Eng Comput. 2002. PMID: 12195978
-
Effect of the starting point of half-pin insertion on the insertional torque of the pin at the tibia.Yonsei Med J. 2015 Jan;56(1):154-9. doi: 10.3349/ymj.2015.56.1.154. Yonsei Med J. 2015. PMID: 25510759 Free PMC article.
-
Enhanced biocompatibility and osseointegration of calcium titanate coating on titanium screws in rabbit femur.J Huazhong Univ Sci Technolog Med Sci. 2017 Jun;37(3):362-370. doi: 10.1007/s11596-017-1741-9. Epub 2017 Jun 6. J Huazhong Univ Sci Technolog Med Sci. 2017. PMID: 28585129
-
Stability with unilateral external fixation in the tibia.Strategies Trauma Limb Reconstr. 2007 Apr;2(1):13-20. doi: 10.1007/s11751-007-0011-y. Strategies Trauma Limb Reconstr. 2007. PMID: 18427910 Free PMC article.
MeSH terms
LinkOut - more resources