Biomechanical analyses of mesh fixation in TAPP and TEP hernia repair
- PMID: 17623239
- DOI: 10.1007/s00464-007-9476-5
Biomechanical analyses of mesh fixation in TAPP and TEP hernia repair
Abstract
Background: Reliable laparoscopic fixation of meshes prior to their fibrous incorporation is intended to minimize recurrences following transabdominal preperitoneal hernia repair (TAPP) and totally extraperitoneal repair (TEP) repair of inguinal hernias. However, suture-, tack- and staple-based fixation systems are associated with postoperative chronic inguinal pain. Initial fixation with fibrin sealant offers an atraumatic alternative, but there is little data demonstrating directly whether fibrin-based mesh adhesion provides adequate biomechanical stability for repair of inguinal hernia by TAPP and TEP.
Methods: Using a newly developed, standardized simulation model for abdominal wall hernias, sublay repairs were performed with six different types of commercially available hernia mesh. The biomechanical stability achieved, and the protection afforded by the mesh-hernia overlap, were compared for three different techniques: nonfixation, point-by-point suture fixation, and fibrin sealant fixation.
Results: Mesh dislocation from the repaired hernia defect was consistently seen with nonfixation. This was reliably prevented with all six mesh types when fixed using either sutures or fibrin sealant. The highest stress resistance across the whole abdominal wall was found following superficial fixation with fibrin sealant across the mesh types. There was a highly statistically significant improvement in fixation stability with fibrin sealant versus fixation using eight single sutures (p = 0.008), as assessed by the range of achievable peak pressure stress up to 200 mmHg.
Conclusions: To ensure long-term freedom from recurrence, intraoperative mesh-hernia overlap must be retained. This can be achieved with fibrin sealant up to the incorporation of the mesh - without trauma and with biomechanical stability.
Similar articles
-
Long-term outcome and chronic pain in atraumatic fibrin glue versus staple fixation of extra light titanized meshes in laparoscopic inguinal hernia repair (TAPP): a single-center experience.Surg Endosc. 2020 May;34(5):1929-1938. doi: 10.1007/s00464-019-06965-x. Epub 2019 Jul 12. Surg Endosc. 2020. PMID: 31300910
-
Review of 1000 fibrin glue mesh fixation during endoscopic totally extraperitoneal (TEP) inguinal hernia repair.Surg Endosc. 2016 Oct;30(10):4544-52. doi: 10.1007/s00464-016-4791-3. Epub 2016 Feb 19. Surg Endosc. 2016. PMID: 26895903
-
Initial experience with the use of fibrin sealant for the fixation of the prosthetic mesh in laparoscopic transabdominal preperitoneal hernia repair.Rozhl Chir. 2005 Aug;84(8):399-402. Rozhl Chir. 2005. PMID: 16218348
-
Intraoperative adjunctive techniques to reduce seroma formation in laparoscopic inguinal hernioplasty: a systematic review.Hernia. 2019 Aug;23(4):723-731. doi: 10.1007/s10029-019-01903-1. Epub 2019 Feb 8. Hernia. 2019. PMID: 30734117
-
Fibrin sealant for mesh fixation in endoscopic inguinal hernia repair: is there enough evidence for its routine use?Surg Laparosc Endosc Percutan Tech. 2010 Aug;20(4):205-12. doi: 10.1097/SLE.0b013e3181ed85b3. Surg Laparosc Endosc Percutan Tech. 2010. PMID: 20729686 Review.
Cited by
-
Study of the biomechanical response of a prosthetic mesh secured with penetrating and non-penetrating fixations in IPOM ventral hernia repair.Hernia. 2024 Feb;28(1):75-84. doi: 10.1007/s10029-023-02915-8. Epub 2023 Nov 14. Hernia. 2024. PMID: 37964087
-
Long-term follow-up of endoscopic totally extraperitoneal direct inguinal hernia repair using the Endoloop technique.Surg Endosc. 2019 Sep;33(9):2967-2974. doi: 10.1007/s00464-018-6602-5. Epub 2018 Nov 26. Surg Endosc. 2019. PMID: 30478697
-
The impact of atraumatic fibrin sealant vs. staple mesh fixation in TAPP hernia repair on chronic pain and quality of life: results of a randomized controlled study.Surg Endosc. 2012 Jan;26(1):249-54. doi: 10.1007/s00464-011-1862-3. Epub 2011 Aug 19. Surg Endosc. 2012. PMID: 21853390 Clinical Trial.
-
Primary mesh augmentation with fibrin glue for abdominal wall closure--investigations on a biomechanical model.Langenbecks Arch Surg. 2010 Feb;395(2):151-6. doi: 10.1007/s00423-008-0453-9. Epub 2009 Jan 22. Langenbecks Arch Surg. 2010. PMID: 19159952
-
A comparison of Progrip(®) and Adhesix (®) self-adhering hernia meshes in an onlay model in the rat.Hernia. 2014 Oct;18(5):761-9. doi: 10.1007/s10029-014-1258-0. Epub 2014 May 6. Hernia. 2014. PMID: 24797638
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources