Supra-Tenon's capsule placement of a single-plate Molteno implant
- PMID: 17627979
- DOI: 10.1136/bjo.2007.115121
Supra-Tenon's capsule placement of a single-plate Molteno implant
Abstract
Aim: To describe clinical results of a new technique for inserting Molteno implants in a supra-Tenon's capsule location.
Design: Prospective interventional case study.
Participants: 19 patients, 18 of whom had suffered previous filter or glaucoma implant failure. Seventeen had open-angle glaucoma, and two had neovascular glaucoma. Seventeen patients were black and two were Asian.
Intervention: All patients underwent supra-Tenon's capsule placement of a single-plate Molteno implant. Goldmann applanation tensions and Snellen visual acuities were determined before the operation and at the last follow-up visit. The numbers of preoperative and postoperative adjunctive drugs were compared before and after surgery.
Main outcome measures: Success was defined as intraocular pressure < or = 18 mm Hg with or without anti-glaucoma drugs.
Results: Mean age was 62 years. Mean pressure before and after surgery was 31 and 16 mm Hg, respectively. Mean follow-up was 22 months (range 12-48). Vision returned to within 2 lines of preoperative vision, except in one patient who lost light perception. Surgery was regarded as a failure in four patients, including both patients with neovascular glaucoma. The difference between the median number of drugs used before (three) and after (one) the operation was significant (p<0.001). Complications were minimal, characterised by small choroidal effusions in three patients; all resolved spontaneously.
Conclusion: Supra-Tenon's capsule placement of Molteno implants in eyes with previously failed glaucoma surgery is an effective method for controlling intraocular pressure.
Similar articles
-
Otago Glaucoma Surgery Outcome Study: follow-up of young patients who underwent Molteno implant surgery.Ophthalmology. 2005 Dec;112(12):2137-42. doi: 10.1016/j.ophtha.2005.07.011. Ophthalmology. 2005. PMID: 16325709
-
[Molteno implant and refractory glaucoma. Evaluation of postoperative IOP control and complications with a modified surgical procedure].J Fr Ophtalmol. 2003 Jan;26(1):15-23. J Fr Ophtalmol. 2003. PMID: 12610405 French.
-
Ahmed glaucoma valve and single-plate Molteno implants in treatment of refractory glaucoma: a comparative study.Am J Ophthalmol. 2010 Jun;149(6):893-902. doi: 10.1016/j.ajo.2010.01.025. Epub 2010 May 8. Am J Ophthalmol. 2010. PMID: 20451896 Clinical Trial.
-
[Problems of modern surgery for glaucoma].Vestn Oftalmol. 2008 Jul-Aug;124(4):53-6. Vestn Oftalmol. 2008. PMID: 18756805 Review. Russian. No abstract available.
-
Management of the late leaking filtration blebs. A report of seven cases and a selective review of the literature.Indian J Ophthalmol. 2001 Dec;49(4):247-54. Indian J Ophthalmol. 2001. PMID: 12930117 Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical