[Sterilization effect analysis of B-class pulsation table top vacuum sterilizer to dental handpieces]
- PMID: 17629204
[Sterilization effect analysis of B-class pulsation table top vacuum sterilizer to dental handpieces]
Abstract
Objective: Discuss sterilization effect of B-class pulsation table top vacuum pressure steam sterilizer for dental handpiece.
Methods: Analysis selection of sterilizer for dental handpiece and sterilization management processes and sterilization effect monitoring, evaluation of monitoring result and effective sterilization method.
Results: The B-class pulsation table top vacuum pressure steam sterilizer to dental handpiece in West China Stomatological Hospital of Sichuan University met the requirement of the chemical and biological monitoring. Its efficiency of sterilization was 100%. The results of aerobic culture, anaerobic culture, B-type hepatitis mark monitoring to sterilized dental handpiece were negative.
Conclusion: It is effective method for dental handpiece sterilization to use B-class pulsation table top vacuum pressure steam sterilizer.
Similar articles
-
Investigating steam penetration using thermometric methods in dental handpieces with narrow internal lumens during sterilizing processes with non-vacuum or vacuum processes.J Hosp Infect. 2017 Dec;97(4):338-342. doi: 10.1016/j.jhin.2017.07.033. Epub 2017 Aug 2. J Hosp Infect. 2017. PMID: 28778810
-
Failure of non-vacuum steam sterilization processes for dental handpieces.J Hosp Infect. 2017 Dec;97(4):343-347. doi: 10.1016/j.jhin.2017.09.004. Epub 2017 Sep 10. J Hosp Infect. 2017. PMID: 28903057
-
Effectiveness of three types of sterilization on the contents of sharps containers.Quintessence Int. 1995 Nov;26(11):771-8. Quintessence Int. 1995. PMID: 8628836
-
Autoclave sterilization of dental handpieces: A literature review.J Prosthodont Res. 2020 Jul;64(3):239-242. doi: 10.1016/j.jpor.2019.07.013. Epub 2019 Aug 12. J Prosthodont Res. 2020. PMID: 31416709 Review.
-
Can oil lubricated dental handpieces be sterilized?: part 1. The problem.Dent Update. 2013 Sep;40(7):543-4, 547-8. doi: 10.12968/denu.2013.40.7.543. Dent Update. 2013. PMID: 24147385 Review.
MeSH terms
LinkOut - more resources
Miscellaneous