Performance of a multislice fan beam collimator for SPECT imaging of the head
- PMID: 1764341
- DOI: 10.1007/BF03164624
Performance of a multislice fan beam collimator for SPECT imaging of the head
Abstract
A multislice fan beam collimator was designed for a SPECT system with a rotating scintillation camera, and it was constructed by the lead casting method which was devised in recent years. The focal length from the surface of the detector side of the collimator is 75 cm. In order to reconstruct the SPECT images we modified the interpolation method which was developed for reconstruction of the X-ray CT and applied it. Primary photopeak data were obtained with a 20% energy window centered at 140 keV for 99mTc and 159 keV for 123I. The fan beam collimator reduces the field of view, and it would therefore be limited to use only for the head and neck region at present; however, both resolution and sensitivity were approximately 20% better than those of the parallel hole collimator. A fan beam collimator is a useful implement for the SPECT study.
Similar articles
-
[Single photon emission computed tomography by using fan beam collimator].Nihon Igaku Hoshasen Gakkai Zasshi. 1992 May 25;52(5):660-8. Nihon Igaku Hoshasen Gakkai Zasshi. 1992. PMID: 1508638 Japanese.
-
Design and clinical utility of a fan beam collimator for SPECT imaging of the head.J Nucl Med. 1986 Jun;27(6):810-9. J Nucl Med. 1986. PMID: 3486963
-
[SPECT images using a multislice fan beam collimator].Radioisotopes. 1986 Oct;35(10):535-9. doi: 10.3769/radioisotopes.35.10_535. Radioisotopes. 1986. PMID: 3492732 Japanese.
-
SPECT instrumentation: performance, lesion detection, and recent innovations.Semin Nucl Med. 1987 Jul;17(3):184-99. doi: 10.1016/s0001-2998(87)80033-8. Semin Nucl Med. 1987. PMID: 3303339 Review.
-
Review of SPECT collimator selection, optimization, and fabrication for clinical and preclinical imaging.Med Phys. 2015 Aug;42(8):4796-813. doi: 10.1118/1.4927061. Med Phys. 2015. PMID: 26233207 Free PMC article. Review.
Cited by
-
Multiple-pinhole collimators improve intra- and between-rater agreement and the certainty of the visual interpretation in dopamine transporter SPECT.EJNMMI Res. 2022 Aug 17;12(1):51. doi: 10.1186/s13550-022-00923-w. EJNMMI Res. 2022. PMID: 35976493 Free PMC article.