Fluorescence rejection in resonance Raman spectroscopy using a picosecond-gated intensified charge-coupled device camera
- PMID: 17650366
- DOI: 10.1366/000370207781269873
Fluorescence rejection in resonance Raman spectroscopy using a picosecond-gated intensified charge-coupled device camera
Abstract
A Raman instrument was assembled and tested that rejects typically 98-99% of background fluorescence. Use is made of short (picosecond) laser pulses and time-gated detection in order to record the Raman signals during the pulse while blocking most of the fluorescence. Our approach uses an ultrafast-gated intensified charge-coupled device (ICCD) camera as a simple and straightforward alternative to ps Kerr gating. The fluorescence rejection efficiency depends mainly on the fluorescence lifetime and on the closing speed of the gate (which is about 80 ps in our setup). A formula to calculate this rejection factor is presented. The gated intensifier can be operated at 80 MHz, so high repetition rates and low pulse energies can be used, thus minimizing photodegradation. For excitation we use a frequency-tripled or -doubled Ti : sapphire laser with a pulse width of 3 ps; it should not be shorter in view of the required spectral resolution. Other critical aspects tested include intensifier efficiency as a function of gate width, uniformity of the gate pulse across the spectrum, and spectral resolution in comparison with ungated detection. The total instrumental resolution is 7 cm(-1) in the blue and 15 cm(-1) in the ultraviolet (UV) region. The setup allows one to use resonance Raman spectroscopy (RRS) for extra sensitivity and selectivity, even in the case of strong background fluorescence. Excitation wavelengths in the visible or UV range no longer have to be avoided. The effectiveness of this setup is demonstrated on a test system: pyrene in the presence of toluene fluorescence (lambda(exc) = 257 nm). Furthermore, good time-gated RRS spectra are shown for a strongly fluorescent flavoprotein (lambda(exc) = 405 nm). Advantages and disadvantages of this approach for RRS are discussed.
Similar articles
-
Steady-state and transient ultraviolet resonance Raman spectrometer for the 193-270 nm spectral region.Appl Spectrosc. 2005 Dec;59(12):1541-52. doi: 10.1366/000370205775142511. Appl Spectrosc. 2005. PMID: 16390595
-
Development of a broadband picosecond infrared spectrometer and its incorporation into an existing ultrafast time-resolved resonance Raman, UV/visible, and fluorescence spectroscopic apparatus.Appl Spectrosc. 2003 Apr;57(4):367-80. doi: 10.1366/00037020360625899. Appl Spectrosc. 2003. PMID: 14658632
-
Fast-gated intensified charge-coupled device camera to record time-resolved fluorescence spectra of tryptophan.Appl Spectrosc. 2004 Jun;58(6):705-10. doi: 10.1366/000370204872908. Appl Spectrosc. 2004. PMID: 15198823
-
Optical system design for biosensors based on CCD detection.Methods Mol Biol. 2009;503:239-58. doi: 10.1007/978-1-60327-567-5_13. Methods Mol Biol. 2009. PMID: 19151945 Review.
-
Achievements in resonance Raman spectroscopy review of a technique with a distinct analytical chemistry potential.Anal Chim Acta. 2008 Jan 14;606(2):119-34. doi: 10.1016/j.aca.2007.11.006. Epub 2007 Nov 9. Anal Chim Acta. 2008. PMID: 18082644 Review.
Cited by
-
Time-resolved Raman spectroscopy using a CMOS SPAD array to remove fluorescent and fibre Raman backgrounds.Biomed Opt Express. 2025 Jun 17;16(7):2824-2834. doi: 10.1364/BOE.560826. eCollection 2025 Jul 1. Biomed Opt Express. 2025. PMID: 40677818 Free PMC article.
-
Spatially offset Raman spectroscopy for biomedical applications.Chem Soc Rev. 2021 Jan 7;50(1):556-568. doi: 10.1039/d0cs00855a. Epub 2020 Nov 10. Chem Soc Rev. 2021. PMID: 33169761 Free PMC article. Review.
-
Insights into Protein Structure and Dynamics by Ultraviolet and Visible Resonance Raman Spectroscopy.Biochemistry. 2015 Aug 11;54(31):4770-83. doi: 10.1021/acs.biochem.5b00514. Epub 2015 Jul 29. Biochemistry. 2015. PMID: 26219819 Free PMC article.
-
Time-Gated Raman Spectroscopy for Quantitative Determination of Solid-State Forms of Fluorescent Pharmaceuticals.Anal Chem. 2018 Apr 3;90(7):4832-4839. doi: 10.1021/acs.analchem.8b00298. Epub 2018 Mar 19. Anal Chem. 2018. PMID: 29513001 Free PMC article.
-
An Effective Baseline Correction Algorithm Using Broad Gaussian Vectors for Chemical Agent Detection with Known Raman Signature Spectra.Sensors (Basel). 2021 Dec 10;21(24):8260. doi: 10.3390/s21248260. Sensors (Basel). 2021. PMID: 34960369 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources