Multicenter evaluation of the BD Phoenix Automated Microbiology System for antimicrobial susceptibility testing of Streptococcus species
- PMID: 17652483
- PMCID: PMC2045298
- DOI: 10.1128/JCM.00981-07
Multicenter evaluation of the BD Phoenix Automated Microbiology System for antimicrobial susceptibility testing of Streptococcus species
Abstract
This multicenter study evaluated the BD Phoenix Automated Microbiology System STREP panel (BD Diagnostic Systems). Antimicrobial susceptibility testing (AST) with 13 agents was performed on 2,013 streptococci (938 Streptococcus pneumoniae isolates; 396 group B streptococci [GBS]; 369 viridans group streptococci [VGS]; 290 beta-hemolytic streptococcus groups A, C, and G; and 20 other streptococci) with the Phoenix system and a broth microdilution reference method. Clinical and challenge isolates were tested against cefepime, cefotaxime (CTX), ceftriaxone (CTR), clindamycin (CLI), erythromycin (ERY), gatifloxacin, levofloxacin, linezolid, meropenem, penicillin (PEN), tetracycline (TET), trimethoprim-sulfamethoxazole, and vancomycin. Clinical isolates with major errors or very major errors (VMEs) were retested in duplicate by both methods. The final results for clinical isolates showed the following trends. For all of the organism-antimicrobial agent combinations tested, categorical agreement (CA) was 92 to 100%, with one exception-VGS-PEN (87% CA; all errors were minor). For S. pneumoniae, there was one major error with CLI (0.1%) and one or two VMEs with CTX (4%), CTR (4.5%), ERY (0.9%), and TET (0.7%). For groups A, C, and G, the CA was 97 to 100% and the only VMEs were resolved by additional reference laboratory testing. For GBS, there was only one VME (TET, 0.3%) and D-zone testing of 23 isolates with CLI major errors (one isolate unavailable) revealed inducible CLI resistance. For VGS, the major error rates were 0 to 3% and VMEs occurred with seven agents (3.5 to 7.1%). The mean times required for organism groups to generate results ranged from 8.4 to 9.4 h. The Phoenix system provided reliable and rapid AST results for most of the organism-antimicrobial agent combinations tested.
References
-
- Beekmann, S. E., K. P. Heilmann, S. S. Richter, J. Garcia-de-Lomas, G. V. Doern, and the GRASP Study Group. 2005. Antimicrobial resistance in Streptococcus pneumoniae, Haemophilus influenzae, Moraxella catarrhalis and group A beta-hemolytic streptococci in 2002-2003; results of the multinational GRASP surveillance program. Int. J. Antimicrob. Agents 25:148-156. - PubMed
-
- Clinical and Laboratory Standards Institute. 2006. Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically; approved standard—seventh edition. CLSI document M7-A7. Clinical and Laboratory Standards Institute, Wayne, PA.
-
- Clinical and Laboratory Standards Institute/NCCLS. 2007. Performance standards for antimicrobial susceptibility testing; 17th informational supplement M100-S17. Clinical and Laboratory Standards Institute, Wayne, PA.
-
- Diekema, D. J., M. L. Beach, M. A. Pfaller, R. N. Jones, and the SENTRY Participant Group. 2001. Antimicrobial resistance in viridans group streptococci among patients with and without the diagnosis of cancer in the USA, Canada, and Latin America. Clin. Microbiol. Infect. 7:152-157. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous