Abstract
In this paper, we present the shielding analysis to determine the necessary neutron and photon shielding for a laser-accelerated proton therapy system. Laser-accelerated protons coming out of a solid high-density target have broad energy and angular spectra leading to dose distributions that cannot be directly used for therapeutic applications. A special particle selection and collimation device is needed to generate desired proton beams for energy- and intensity-modulated proton therapy. A great number of unwanted protons and even more electrons as a side-product of laser acceleration have to be stopped by collimation devices and shielding walls, posing a challenge in radiation shielding. Parameters of primary particles resulting from the laser-target interaction have been investigated by particle-in-cell simulations, which predicted energy spectra with 300 MeV maximum energy for protons and 270 MeV for electrons at a laser intensity of 2 x 10(21) W cm(-2). Monte Carlo simulations using FLUKA have been performed to design the collimators and shielding walls inside the treatment gantry, which consist of stainless steel, tungsten, polyethylene and lead. A composite primary collimator was designed to effectively reduce high-energy neutron production since their highly penetrating nature makes shielding very difficult. The necessary shielding for the treatment gantry was carefully studied to meet the criteria of head leakage <0.1% of therapeutic absorbed dose. A layer of polyethylene enclosing the whole particle selection and collimation device was used to shield neutrons and an outer layer of lead was used to reduce photon dose from neutron capture and electron bremsstrahlung. It is shown that the two-layer shielding design with 10-12 cm thick polyethylene and 4 cm thick lead can effectively absorb the unwanted particles to meet the shielding requirements.
Similar articles
-
Dosimetric advantages of IMPT over IMRT for laser-accelerated proton beams.Phys Med Biol. 2008 Dec 21;53(24):7151-66. doi: 10.1088/0031-9155/53/24/010. Epub 2008 Nov 26. Phys Med Biol. 2008. PMID: 19033641
-
Bremsstrahlung and photoneutron production in a steel shield for 15-22-MeV clinical electron beams.Radiat Prot Dosimetry. 2015 Feb;163(2):148-59. doi: 10.1093/rpd/ncu153. Epub 2014 May 12. Radiat Prot Dosimetry. 2015. PMID: 24821930
-
Dual scattering foil design for poly-energetic electron beams.Phys Med Biol. 2005 Mar 7;50(5):755-67. doi: 10.1088/0031-9155/50/5/002. Epub 2005 Feb 17. Phys Med Biol. 2005. PMID: 15798252
-
Dosimetry of clinical neutron and proton beams: an overview of recommendations.Radiat Prot Dosimetry. 2004;110(1-4):565-72. doi: 10.1093/rpd/nch221. Radiat Prot Dosimetry. 2004. PMID: 15353710 Review.
-
Influence of radiotherapy room shielding on ambient dose equivalent due to photons H*(10)p and neutrons H*(10)n in the patient's plane.Appl Radiat Isot. 2022 Mar;181:110095. doi: 10.1016/j.apradiso.2021.110095. Epub 2022 Jan 5. Appl Radiat Isot. 2022. PMID: 34999307 Review.
Cited by
-
Geometrical splitting technique to improve the computational efficiency in Monte Carlo calculations for proton therapy.Med Phys. 2013 Apr;40(4):041718. doi: 10.1118/1.4795343. Med Phys. 2013. PMID: 23556888 Free PMC article.
-
Monte Carlo Simulations of Neutron Ambient Dose Equivalent in a Novel Proton Therapy Facility Design.Int J Part Ther. 2020 Spring;6(4):29-37. doi: 10.14338/IJPT-19-00071.1. Epub 2020 Mar 12. Int J Part Ther. 2020. PMID: 32582817 Free PMC article.
-
Geant4 Monte Carlo simulation study of the secondary radiation fields at the laser-driven ion source LION.Sci Rep. 2021 Dec 24;11(1):24418. doi: 10.1038/s41598-021-03897-2. Sci Rep. 2021. PMID: 34952912 Free PMC article.
-
Improved efficiency in Monte Carlo simulation for passive-scattering proton therapy.Phys Med Biol. 2015 Jul 7;60(13):5019-35. doi: 10.1088/0031-9155/60/13/5019. Epub 2015 Jun 10. Phys Med Biol. 2015. PMID: 26061457 Free PMC article.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources