Variability of repeated coronary artery calcium measurements by 1.25-mm- and 2.5-mm-thickness images on prospective electrocardiograph-triggered 64-slice CT

Affiliations
Controlled Clinical Trial

Variability of repeated coronary artery calcium measurements by 1.25-mm- and 2.5-mm-thickness images on prospective electrocardiograph-triggered 64-slice CT

Jun Horiguchi et al. Eur Radiol. 2008 Feb.

Abstract

High reproducibility on coronary artery calcium scoring is a key requirement in monitoring the progression of coronary atherosclerosis. The purpose of this prospective study is to assess the reproducibility of 1.25-mm- and 2.5-mm-thickness images on prospective electrocardiograph-triggered 64-slice CT with respect to 2.5-mm-thickness images on spiral overlapping reconstruction. One hundred patients suspected of coronary artery disease were scanned twice repeatedly, both on prospective electrocardiograph-triggered step-and-shoot and retrospective electrocardiograph-gated spiral scans. Using 1.25-mm-thickness collimation, 1.25-mm- and 2.5-mm-thickness image sets on prospective scans and 2.5-mm-thickness image sets with 1.25-mm increment (overlapping) on retrospective scans were obtained. Coronary artery calcium scores, interscan variability and interobserver variability were evaluated. The mean interscan variability in coronary artery calcium measurement on 1.25-mm prospective/2.5-mm prospective/2.5-mm overlapping retrospective scans were Agatston: 10%/18%/12%, volume: 10%/12%/10% and mass: 8%/13%/11% for observer 1 and Agatston: 8%/14%/10%, volume: 7%/9%/10% and mass: 7%/10%/9% for observer 2, respectively. The mean interobserver variability was 5% to 14%. In conclusion, prospective electrocardiograph-triggered 64-slice CT using the 1.25-mm prospective scan shows the lowest variability. The 2.5-mm prospective scan on volume or mass scoring shows variability of around 10%, comparable to 2.5-mm-thickness spiral overlapping reconstruction images.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Comput Assist Tomogr. 2001 Jul-Aug;25(4):569-73 - PubMed
    1. Circulation. 2003 Feb 18;107(6):917-22 - PubMed
    1. Acad Radiol. 2006 Feb;13(2):173-6 - PubMed
    1. Mayo Clin Proc. 1999 Apr;74(4):347-55 - PubMed
    1. Am J Cardiol. 1991 Jul 1;68(1):1-6 - PubMed

Publication types