Comparative Study
. 2007;3(3):343-9.

Noninvasive measurements of arterial stiffness: repeatability and interrelationships with endothelial function and arterial morphology measures

Affiliations
Comparative Study

Noninvasive measurements of arterial stiffness: repeatability and interrelationships with endothelial function and arterial morphology measures

Corey J Huck et al. Vasc Health Risk Manag. 2007.

Abstract

Background: Many noninvasive arterial assessment techniques have been developed, measuring different parameters of arterial stiffness and endothelial function. However, there is little data available comparing different devices within the same subject. Therefore, the purpose of this study was to examine the repeatability and interrelationships between 3 different techniques to measure arterial stiffness and to compare this with forearm-mediated dilation.

Methods: Carotid-radial pulse wave velocity was measured by the Sphygmocor (SPWV) and Complior (CPWV) devices, cardio-ankle vascular index (CAVI) was measured by the VaSera device, vascular structure and function was assessed using ultrasonography and evaluated for reliability and compared in 20 apparently healthy, college-aged men and women.

Results: The intraclass correlation coefficient and standard error of the mean for the Sphygmocor (R = 0.56, SEM = 0.69), Complior (R = 0.62, SEM = 0.69), and VaSera (R = 0.60, SEM = 0.56), indicated moderate repeatability. Bland-Altman plots indicated a mean difference of 0.11 +/- 0.84 for SPWV, 0.13 +/- 1.15 for CPWV, and -0.43 +/- 0.90 for CAVI. No significant interrelationships were found among the ultrasound measures and SPWV, CPWV, and CAVI.

Conclusions: The three noninvasive modalities to study arterial stiffness reliably measures arterial stiffness however, they do not correlate with ultrasound measures of vascular function and structure in young and apparently healthy subjects.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Bland-Altman plot showing the intra-observer differences against the mean Pulse wave velocity value of Sphygomocor (Panel A), Complior (Panel B), and Cardiac-ankle index VaSera (Panel C). Dotted line represents two standard deviations from the mean value.

Similar articles

Cited by

References

    1. Asmar R, Benetos A, Topouchian J, et al. Assessment of arterial distensibility by automatic pulse wave velocity measurement. Validation and Clinical application studies. Hypertension. 1995;26:485–90. - PubMed
    1. Asmar R, Topouchian J, Pannier B, et al. Pulse wave velocity as endpoint in large-scale intervention trial (The Complior Study) J Hypertens. 2001;19:813–18. - PubMed
    1. Avest ET, Holewijn S, Stalenhoef AFH, et al. Variation in non-invasive measurements of vascular function in healthy volunteers during daytime. Clin Science. 2005;108:425–31. - PubMed
    1. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;i:307–10. - PubMed
    1. Boutouyrie P, Bussy C, Lacolley P, et al. Association between local pulse pressure, mean blood pressure, and large-artery remodeling. Circ. 1999;100:1387–93. - PubMed