The accuracy of root canal measurements using the Mini Apex Locator and Root ZX-II: an evaluation in vitro
- PMID: 17709069
- DOI: 10.1016/j.tripleo.2007.03.025
The accuracy of root canal measurements using the Mini Apex Locator and Root ZX-II: an evaluation in vitro
Abstract
Objective: The aim of this in vitro study was to compare the capacity of the Root ZX-II and Mini Apex Locator, electronic apex locators (EALS), to prevent overestimated working length.
Study design: Forty extracted human teeth were used for the study. The cervical portion of each canal was flared using Gates Glidden drills and the teeth were embedded in an alginate model. Canals were irrigated with 2.5% sodium hypochlorite. The actual length (AL) and electronic length (EL) measurements were made on each specimen separately with both devices with the aid of a k-type file.
Results: The results obtained with each EAL were compared with the corresponding actual length. The statistical analysis of the results showed EAL reliability to prevent overestimated working length to be 100% for the Mini Apex Locator and 97.44% for the Root ZX-II, within a tolerance of +/-0.5 mm into account. A paired sample t test showed that there was no statistically significant difference between the accuracy of the devices (P = .5841).
Conclusion: The results of this study indicate that the Root ZX-II and Mini Apex Locator are accurate devices to prevent overestimated working length.
Similar articles
-
The ability of two apex locators to locate the apical foramen: an in vitro study.J Endod. 2006 Jun;32(6):560-2. doi: 10.1016/j.joen.2005.11.011. Epub 2006 Apr 4. J Endod. 2006. PMID: 16728251
-
Accuracy of a new apex locator: an in vitro study.Int Endod J. 2002 Feb;35(2):186-92. doi: 10.1046/j.1365-2591.2002.00468.x. Int Endod J. 2002. PMID: 11843975
-
Influence of preflaring on the accuracy of length determination with four electronic apex locators.J Endod. 2009 Sep;35(9):1300-2. doi: 10.1016/j.joen.2009.05.030. J Endod. 2009. PMID: 19720236
-
Electronic apex locator.Dent Clin North Am. 2004 Jan;48(1):35-54. doi: 10.1016/j.cden.2003.10.005. Dent Clin North Am. 2004. PMID: 15066506 Review.
-
[Determination of canal length in endodontics].Rev Belge Med Dent (1984). 2005;60(1):31-40. Rev Belge Med Dent (1984). 2005. PMID: 15943057 Review. French.
Cited by
-
Root ZX Electronic Foramen Locator: An Ex Vivo Study of Its Three Models' Precision and Reproducibility.Int J Dent. 2017;2017:5893790. doi: 10.1155/2017/5893790. Epub 2017 Mar 7. Int J Dent. 2017. PMID: 28367215 Free PMC article.
-
Efficiency of 2 electronic apex locators on working length determination: A clinical study.J Conserv Dent. 2013 May;16(3):229-32. doi: 10.4103/0972-0707.111320. J Conserv Dent. 2013. PMID: 23833456 Free PMC article.
-
An in vivo comparison of accuracy of two electronic apex locators in determining working length using stainless steel and nickel titanium files.Clin Cosmet Investig Dent. 2018 May 22;10:75-82. doi: 10.2147/CCIDE.S158882. eCollection 2018. Clin Cosmet Investig Dent. 2018. PMID: 29861643 Free PMC article.
-
Comparison of accuracy of two electronic apex locators in the presence of various irrigants: An in vitro study.J Conserv Dent. 2012 Apr;15(2):178-82. doi: 10.4103/0972-0707.94585. J Conserv Dent. 2012. PMID: 22557820 Free PMC article.
-
In vitro comparison of working length determination using three different electronic apex locators.Dent Res J (Isfahan). 2014 Sep;11(5):568-73. Dent Res J (Isfahan). 2014. PMID: 25426148 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous