Tip position of long-term central venous access devices used for parenteral nutrition
- PMID: 17712146
- DOI: 10.1177/0148607107031005382
Tip position of long-term central venous access devices used for parenteral nutrition
Abstract
Background: Venous thrombosis is a potential postplacement complication of a central venous access device (VAD). Improper catheter tip position is a predisposing factor, especially when the device is used to administer parenteral nutrition (PN). American Society for Parenteral and Enteral Nutrition (A.S.P.E.N.) guidelines recommend that a central VAD used for PN be placed with its tip in the superior vena cava (SVC) adjacent to the right atrium (RA). The purpose of this study is to determine the prevalence of improper central VAD tip position and factors associated with malpositioning.
Methods: All adult patients with a longterm VAD (ie, tunneled central venous catheter, peripherally inserted central catheter [PICC], or implanted port) placed before the current admission who were scheduled to receive PN also received chest x-rays to evaluate position of the catheter tip. Position was determined by a staff radiologist. A catheter with its tip ranging from the middle third of the SVC to the RA was considered acceptable; a catheter with its tip in any other position was considered malpositioned. Subjects with multiple VADs or multiple evaluations for the same catheter had the first placement and last evaluation considered. A logistic regression analysis was used to study the univariable and multivariable associations of these factors with tip malposition.
Results: Data were collected for catheters in 124 patients, including 74 tunneled catheters (71 Hickman, 2 Broviac, 1 Groshong), 38 PICCs, and implanted ports. Most of the catheters were placed for (81.9%) or chemotherapy (14.5%). Median catheter duration was 1.6 months at time of evaluation. Of 138 catheters studied, 15.9% (95% confidence interval, 10.2-23.1) were malpositioned at time of evaluation. According to univariable analysis, factors associated with malpositioned catheters included shorter catheter duration (p = .001), greater number of lumens (p = .029), venous entry site on the arm (p <.001) and catheters placed at institutions other than Cleveland Clinic (p = .007). Additionally, PICCs were likely to be malpositioned at time of evaluation compared with other long-term VADs combined (34.2% vs 9.0%; p < .001).
Conclusions: A high percentage of long-term VADs improperly positioned for PN in the present study. were more likely to be malpositioned at time of evaluation compared with tunneled catheters and implanted These findings suggest the tip position of long-term should be confirmed before infusing PN.
Similar articles
-
ESPEN Guidelines on Parenteral Nutrition: central venous catheters (access, care, diagnosis and therapy of complications).Clin Nutr. 2009 Aug;28(4):365-77. doi: 10.1016/j.clnu.2009.03.015. Epub 2009 May 21. Clin Nutr. 2009. PMID: 19464090
-
A comparative study of peripherally-inserted and Broviac catheter complications in home parenteral nutrition patients.Clin Nutr. 2015 Feb;34(1):49-52. doi: 10.1016/j.clnu.2013.12.017. Epub 2014 Jan 3. Clin Nutr. 2015. PMID: 24439240
-
Peripherally inserted central catheters for long-term parenteral nutrition in infants with intestinal failure.J Pediatr Gastroenterol Nutr. 2013 May;56(5):578-81. doi: 10.1097/MPG.0b013e3182801e7b. J Pediatr Gastroenterol Nutr. 2013. PMID: 23221995
-
Principles of tunneled cuffed catheter placement.Tech Vasc Interv Radiol. 2011 Dec;14(4):192-7. doi: 10.1053/j.tvir.2011.05.008. Tech Vasc Interv Radiol. 2011. PMID: 22099010 Review.
-
Subcutaneous infusion ports for administration of parenteral nutrition at home.Surg Gynecol Obstet. 1989 Oct;169(4):329-33. Surg Gynecol Obstet. 1989. PMID: 2506656 Review.
Cited by
-
Prevention of catheter tip malposition with an ultrasound-guided finger-pressure method to block the internal jugular vein during PICC placement: a meta-analysis.Wideochir Inne Tech Maloinwazyjne. 2022 Jun;17(2):289-298. doi: 10.5114/wiitm.2022.115198. Epub 2022 Apr 4. Wideochir Inne Tech Maloinwazyjne. 2022. PMID: 35707337 Free PMC article.
-
Chylothorax associated with thrombosis of the cranial vena cava.Can Vet J. 2010 Aug;51(8):847-52. Can Vet J. 2010. PMID: 21037884 Free PMC article.
-
The utilization of guidewires for adjusting the intraoperative catheter malposition during the venous access port implantation: A retrospective study.Medicine (Baltimore). 2024 Nov 8;103(45):e40461. doi: 10.1097/MD.0000000000040461. Medicine (Baltimore). 2024. PMID: 39533627 Free PMC article.
-
Outcome analysis in 3,160 implantations of radiologically guided placements of totally implantable central venous port systems.Eur Radiol. 2011 Jun;21(6):1224-32. doi: 10.1007/s00330-010-2045-7. Epub 2011 Jan 5. Eur Radiol. 2011. PMID: 21207035
-