[Image-guided minimal-invasive cochlear implantation--experiments on cadavers]
- PMID: 17713878
- DOI: 10.1055/s-2007-966775
[Image-guided minimal-invasive cochlear implantation--experiments on cadavers]
Abstract
Background: The accuracy of navigation systems can be improved significantly by using high-resolution flat panel-based Volume Computed Tomography (fpVCT) so that new surgical therapeutic concepts become feasible. A navigation-guided minimally-invasive cochleostomy places highest requirements on the accuracy of intraoperative navigation.
Methods: A flat-panel Volume Computed Tomograph (fpVCT) was used to scan four human temporal bones. The isometric voxel size was 200 microm. The preoperative planning was used to define an optimized drilling channel from the mastoid surface to the round window niche and the scala tympani providing a safety margin to critical anatomical structures such as facial nerve, chorda tympani, sigmoid sinus and posterior wall of auditory canal. The canal was drilled hand-operated with a navigated drill following the previously planned trajectory. Afterwards the drilled canal was imaged by fpVCT. Conventional dissection including mastoidectomy and posterior tympanotomy assured correct localization of the cochleostomy.
Results: Path planning took an average of 54 minutes (range 35-85 minutes). Installation took an average of 16 minutes (range 14-19 minutes). The drilling procedure itself took an average of 7.75 min (range 5-12 minutes.) The RMSE-values varied between 0.1 and 0.2 mm (Table 1). All four specimens showed a cochleostomy located at the scala tympani anterior inferior to the round window. The chorda tympani was damaged in one specimen--this was preoperatively planned as a narrow facial recess was encountered. The time needed for planning and system-installation could be reduced continuously.
Conclusions: This feasibility study demonstrates that using current image-guided surgery technology in combination with fpVCT allows drilling of a minimally invasive channel to the cochlea with loco typico cochleostomy. The necessary accuracy of intraoperative navigation can be achieved by use of fpVCT (technical accuracy between 0.1 and 0.2 mm). Our results demonstrate the feasibility of a navigation-guided minimally-invasive cochleostomy loco typico. While we are enthused by this preliminary work, we recognize the barriers which exist in translation to clinical application. These include surgical issues (e.g. control of unexpected bleeding) and electrode issues (e.g. development of insertion tools).
Similar articles
-
A true minimally invasive approach for cochlear implantation: high accuracy in cranial base navigation through flat-panel-based volume computed tomography.Otol Neurotol. 2008 Feb;29(2):120-3. doi: 10.1097/mao.0b013e318157f7d8. Otol Neurotol. 2008. PMID: 17971720
-
In vitro accuracy evaluation of image-guided robot system for direct cochlear access.Otol Neurotol. 2013 Sep;34(7):1284-90. doi: 10.1097/MAO.0b013e31829561b6. Otol Neurotol. 2013. PMID: 23921934
-
Image guided navigation by intraoperative CT scan for cochlear implantation.Comput Aided Surg. 2012;17(3):153-60. doi: 10.3109/10929088.2012.668937. Epub 2012 Mar 27. Comput Aided Surg. 2012. PMID: 22452534
-
Image-guided technique in neurotology.Otolaryngol Clin North Am. 2007 Jun;40(3):611-24, x. doi: 10.1016/j.otc.2007.03.006. Otolaryngol Clin North Am. 2007. PMID: 17544698 Review.
-
Workflow and simulation of image-to-physical registration of holes inside spongy bone.Int J Comput Assist Radiol Surg. 2017 Aug;12(8):1425-1437. doi: 10.1007/s11548-017-1594-5. Epub 2017 May 6. Int J Comput Assist Radiol Surg. 2017. PMID: 28478518 Review.
Cited by
-
Automated insertion of preformed cochlear implant electrodes: evaluation of curling behaviour and insertion forces on an artificial cochlear model.Int J Comput Assist Radiol Surg. 2010 Mar;5(2):173-81. doi: 10.1007/s11548-009-0299-9. Epub 2009 Apr 10. Int J Comput Assist Radiol Surg. 2010. PMID: 20033522
-
A robot-guided minimally invasive approach for cochlear implant surgery: preliminary results of a temporal bone study.Int J Comput Assist Radiol Surg. 2009 Sep;4(5):475-86. doi: 10.1007/s11548-009-0360-8. Epub 2009 Jun 13. Int J Comput Assist Radiol Surg. 2009. PMID: 20033531
-
Determination of the curling behavior of a preformed cochlear implant electrode array.Int J Comput Assist Radiol Surg. 2011 May;6(3):421-33. doi: 10.1007/s11548-010-0520-x. Epub 2010 Jul 28. Int J Comput Assist Radiol Surg. 2011. PMID: 20665247
-
Highly precise navigation at the lateral skull base by the combination of flat-panel volume CT and electromagnetic navigation.Sci Prog. 2021 Jul-Sep;104(3):368504211032090. doi: 10.1177/00368504211032090. Sci Prog. 2021. PMID: 34397283 Free PMC article.
-
[Soft tissue navigation and image-guided removal of foreign bodies in head and neck surgery].HNO. 2009 Oct;57(10):1016-22. doi: 10.1007/s00106-008-1779-x. HNO. 2009. PMID: 18688586 Clinical Trial. German.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources