Limitations of arterial pulse pressure variation and left ventricular stroke volume variation in estimating cardiac pre-load during open heart surgery
- PMID: 17714575
- DOI: 10.1111/j.1399-6576.2007.01423.x
Limitations of arterial pulse pressure variation and left ventricular stroke volume variation in estimating cardiac pre-load during open heart surgery
Abstract
Background: In addition to their well-known ability to predict fluid responsiveness, functional pre-load parameters, such as the left ventricular stroke volume variation (SVV) and pulse pressure variation (PPV), have been proposed to allow real-time monitoring of cardiac pre-load. SVV and PPV result from complex heart-lung interactions during mechanical ventilation. It was hypothesized that, under open-chest conditions, when cyclic changes in pleural pressures during positive-pressure ventilation are less pronounced, functional pre-load indicators may be deceptive in the estimation of ventricular pre-load.
Methods: Forty-five patients undergoing coronary artery bypass grafting participated in this prospective, observational study. PPV and SVV were assessed by pulse contour analysis. The thermodilution technique was used to measure the stroke volume index and global and right ventricular end-diastolic volume index. Trans-oesophageal echocardiography was used to determine the left ventricular end-diastolic area index. All parameters were assessed before and after sternotomy, and, in addition, after weaning from cardiopulmonary bypass before and after chest closure (pericardium left open). Patients were ventilated with constant tidal volumes (8 +/- 2 ml/kg) throughout the study period using pressure control.
Results: SVV and PPV decreased after sternotomy and increased after chest closure. However, these changes could not be related to concomitant changes in the ventricular pre-load. The stroke volume index was correlated with SVV and PPV in closed-chest conditions only, whereas volumetric indices reflected cardiac pre-load in both closed- and open-chest conditions. SVV and PPV were correlated with left and right ventricular pre-load in closed-chest-closed-pericardium conditions only (with the best correlation found for the right ventricular end-diastolic volume index).
Conclusions: SVV and PPV may be misleading when estimating cardiac pre-load during open heart surgery.
Similar articles
-
Dynamic preload indicators fail to predict fluid responsiveness in open-chest conditions.Crit Care Med. 2009 Feb;37(2):510-5. doi: 10.1097/CCM.0b013e3181958bf7. Crit Care Med. 2009. PMID: 19114886
-
Effects of mid-line thoracotomy on the interaction between mechanical ventilation and cardiac filling during cardiac surgery.Br J Anaesth. 2004 Jun;92(6):808-13. doi: 10.1093/bja/aeh151. Epub 2004 Apr 19. Br J Anaesth. 2004. PMID: 15096443
-
Effects of sternotomy on heart-lung interaction in patients undergoing cardiac surgery receiving pressure-controlled mechanical ventilation.Acta Anaesthesiol Scand. 2007 Apr;51(4):441-6. doi: 10.1111/j.1399-6576.2007.01245.x. Acta Anaesthesiol Scand. 2007. PMID: 17378782
-
Monitoring fluid responsiveness.Acta Anaesthesiol Taiwan. 2011 Jun;49(2):59-65. doi: 10.1016/j.aat.2011.05.001. Epub 2011 Jun 24. Acta Anaesthesiol Taiwan. 2011. PMID: 21729812 Review.
-
Emerging trends in minimally invasive haemodynamic monitoring and optimization of fluid therapy.Eur J Anaesthesiol. 2009 Nov;26(11):893-905. doi: 10.1097/EJA.0b013e3283308e50. Eur J Anaesthesiol. 2009. PMID: 19667998 Review.
Cited by
-
Predicting fluid responsiveness in 100 critically ill children: the effect of baseline contractility.Intensive Care Med. 2015 Dec;41(12):2161-9. doi: 10.1007/s00134-015-4075-8. Epub 2015 Sep 28. Intensive Care Med. 2015. PMID: 26415680
-
Dynamic and volumetric variables reliably predict fluid responsiveness in a porcine model with pleural effusion.PLoS One. 2013;8(2):e56267. doi: 10.1371/journal.pone.0056267. Epub 2013 Feb 13. PLoS One. 2013. PMID: 23418546 Free PMC article.
-
Stroke volume variation for predicting responsiveness to fluid therapy in patients undergoing cardiac and thoracic surgery: a systematic review and meta-analysis.BMJ Open. 2022 May 18;12(5):e051112. doi: 10.1136/bmjopen-2021-051112. BMJ Open. 2022. PMID: 35584881 Free PMC article.
-
A systematic review of pulse pressure variation and stroke volume variation to predict fluid responsiveness during cardiac and thoracic surgery.J Clin Monit Comput. 2017 Aug;31(4):677-684. doi: 10.1007/s10877-016-9898-5. Epub 2016 Jun 15. J Clin Monit Comput. 2017. PMID: 27306799
-
Does pulse pressure variation predict fluid responsiveness in critically ill patients? A systematic review and meta-analysis.Crit Care. 2014 Nov 27;18(6):650. doi: 10.1186/s13054-014-0650-6. Crit Care. 2014. PMID: 25427970 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical