. 2007 Oct;45(10):3436-8.
doi: 10.1128/JCM.00973-07. Epub 2007 Aug 22.

Human immunodeficiency virus type 1 (HIV-1) plasma load discrepancies between the Roche COBAS AMPLICOR HIV-1 MONITOR Version 1.5 and the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 assays

Affiliations

Human immunodeficiency virus type 1 (HIV-1) plasma load discrepancies between the Roche COBAS AMPLICOR HIV-1 MONITOR Version 1.5 and the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 assays

F Damond et al. J Clin Microbiol. 2007 Oct.

Abstract

We compared plasma viral load values obtained with COBAS AMPLICOR human immunodeficiency virus type 1 (HIV-1) MONITOR version 1.5 and with COBAS TaqMan HIV-1 assays. Mean values were 4.2 and 2.9 log(10) copies/ml, respectively, showing the lack of agreement between the two assays.

PubMed Disclaimer

Figures

FIG. 1.
FIG. 1.
Degree of agreement in log10 copies/ml between the COBAS MONITOR 1.5 and the COBAS TaqMan assays for the 160 plasma specimens. The x axis of the Bland and Altman curves bears the mean values for each sample obtained by the two techniques. The y axis bears the differences between the values obtained by the two techniques. The solid lines represent the mean differences between the values, and the dotted lines represent the mean difference plus or minus 1.96 SD (95% limit of agreement).

Comment in

References

    1. Alaeus, A., K. Lidman, A. Sonnerborg, and J. Albert. 1997. Subtype-specific problems with quantification of plasma HIV-1 RNA. AIDS 11:859-865. - PubMed
    1. Berger, A., L. Scherzed, M. Sturmer, W. Preiser, H. W. Doerr, and H. F. Rabenau. 2002. Evaluation of the COBAS AmpliPrep/Cobas Amplicor HIV-1 Monitor Ultrasensitive Test: comparison with the Cobas Amplicor HIV-1 Monitor test (manual specimen preparation). J. Clin. Virol. 25(Suppl. 3):S103-S107. - PubMed
    1. Damond, F., C. Apetrei, D. Descamps, F. Brun-Vezinet, and F. Simon. 1999. HIV-1 subtypes and plasma RNA quantification. AIDS 13:286-288. - PubMed
    1. Gueudin, M., J. C. Plantier, V. Lemee, M. P. Schmitt, L. Chartier, T. Bourlet, A. Ruffault, F. Damond, M. Vray, and F. Simon. 2007. Evaluation of the Roche Cobas TaqMan and Abbott RealTime Extraction-Quantification systems for HIV-1 subtypes. J. Acquir. Immune Defic. Syndr. 44:500-505. - PubMed
    1. Mellors, J. W., A. Munoz, J. V. Giorgi, J. B. Margolick, C. J. Tassoni, P. Gupta, L. A. Kingsley, J. A. Todd, A. J. Saah, R. Detels, J. P. Phair, and C. R. Rinaldo, Jr. 1997. Plasma viral load and CD4+ lymphocytes as prognostic markers of HIV-1 infection. Ann. Intern. Med. 126:946-954. - PubMed

LinkOut - more resources