Resting energy expenditure and thermal balance during isothermic and thermoneutral haemodialysis--heat production does not explain increased body temperature during haemodialysis
- PMID: 17717032
- DOI: 10.1093/ndt/gfm436
Resting energy expenditure and thermal balance during isothermic and thermoneutral haemodialysis--heat production does not explain increased body temperature during haemodialysis
Abstract
Background: During routine haemodialysis (HD) body temperature increases, which contributes to haemodynamic instability. The relative roles of increased heat production and/or incomplete heat transfer are not fully elucidated. Concomitant measurement of heat production and heat transfer may help to assess the factors determining thermal balance during HD.
Methods: Thirteen stable non-diabetic maintenance HD patients were investigated during two HD procedures (isothermic, dT = 0, no change of body temperature; thermoneutral, dE = 0, no energy transfer between blood and dialysate), using a blood temperature monitor (BTM) in active mode. Energy transfer, blood and dialysate temperature, and relative blood volume change (dBV) were continuously recorded, and resting energy expenditure (REE; Deltatrac Datex) was measured repeatedly during each procedure. Fourteen healthy persons served as controls for REE comparison.
Results: In isothermic HD, median energy removal was 218 kJ/4 h HD (= heat flow -15.1 W). This cooling correlated with dBV induced by ultrafiltration (rho = 0.731, P < 0.01). There was no difference in dBV between isothermic (7.7%) and thermoneutral (8.1%) HD. Predialysis REE was 82.8 W/1.73 m(2), not different from controls. No variation in REE during HD was observed, except a small and transient increase after a light meal (5 and 4%). In the time course of REE, no difference between the procedures was found.
Conclusions: Our findings suggest that stable maintenance HD patients have REE not different from healthy controls, that HD procedure per se does not significantly increase REE and that neither isothermic nor thermoneutral regimen has any influence on metabolic rate. Therefore, body temperature elevation during routine HD may rather be due to decreased heat removal. With the use of BTM in active mode, body temperature can be kept stable (isothermic HD), which requires active cooling. This negative energy transfer is proportional to decrease in blood volume induced by ultrafiltration.
Similar articles
-
[Resting energy expenditure during hemodialysis].Vnitr Lek. 2006 Jan;52(1):26-33. Vnitr Lek. 2006. PMID: 16526195 Czech.
-
Effect of ultrafiltration on thermal variables, skin temperature, skin blood flow, and energy expenditure during ultrapure hemodialysis.J Am Soc Nephrol. 2005 Jun;16(6):1824-31. doi: 10.1681/ASN.2004080655. Epub 2005 Apr 27. J Am Soc Nephrol. 2005. PMID: 15857923
-
Isothermic hemodialysis and ultrafiltration.Am J Kidney Dis. 2000 Aug;36(2):353-61. doi: 10.1053/ajkd.2000.8986. Am J Kidney Dis. 2000. PMID: 10922314
-
[From dialysate temperature to thermal balance].G Ital Nefrol. 2006 Jan-Feb;23(1):29-36. G Ital Nefrol. 2006. PMID: 16521073 Review. Italian.
-
Haemodialysis with on-line monitoring equipment: tools or toys?Nephrol Dial Transplant. 2005 Jan;20(1):22-33. doi: 10.1093/ndt/gfh555. Nephrol Dial Transplant. 2005. PMID: 15632348 Review.
Cited by
-
Feedback control in hemodialysis.Semin Dial. 2025 Jan-Feb;38(1):62-70. doi: 10.1111/sdi.13185. Epub 2023 Nov 23. Semin Dial. 2025. PMID: 37994191 Free PMC article. Review.
-
Lower Dialysate Temperature in Hemodialysis: Is It a Cool Idea?Clin J Am Soc Nephrol. 2015 Aug 7;10(8):1318-20. doi: 10.2215/CJN.06920615. Epub 2015 Jul 20. Clin J Am Soc Nephrol. 2015. PMID: 26195504 Free PMC article. No abstract available.
-
Hemodialysis-Nutritional Flaws in Diagnosis and Prescriptions. Could Amino Acid Losses be the Sharpest "Sword of Damocles"?Nutrients. 2020 Jun 14;12(6):1773. doi: 10.3390/nu12061773. Nutrients. 2020. PMID: 32545868 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical