Dynamic analysis of dark-light changes of the anterior chamber angle with anterior segment OCT
- PMID: 17724195
- DOI: 10.1167/iovs.07-0010
Dynamic analysis of dark-light changes of the anterior chamber angle with anterior segment OCT
Abstract
Purpose: To describe the use of anterior segment optical coherence tomography (OCT) in studying the dynamic dark-light changes of the anterior chamber angle.
Methods: Thirty-seven normal subjects with open angles on dark-room gonioscopy and 18 subjects with narrow angles were analyzed. The dynamic dark-light changes of the anterior-chamber angle were captured with real-time video recording. The angle opening distance (AOD500) and trabecular iris space area (TISA500) of the nasal angle and the pupil diameter in each of the representative serial images were measured. Linear regression analysis was performed to investigate the association between AOD500/TISA500 and pupil diameter. Demographic and biometry measurements associated with the AOD difference (AOD500((light)) - AOD500((dark))) and TISA difference (TISA500((light)) - TISA500((dark))) were analyzed with univariate and multivariate regression models.
Results: The AOD500/TISA500 measured in the light in the open-angle and the narrow-angle groups were 694 +/- 330 microm/0.24 +/- 0.10 mm(2) and 265 +/- 78 microm/0.10 +/- 0.03 mm(2), respectively. These values were significantly greater than the AOD500/TISA500 measured in the dark (492 +/- 265 microm/0.16 +/- 0.08 mm(2) and 119 +/- 82 microm/0.05 +/- 0.04 mm(2), respectively, all with P < 0.001). The ranges of the AOD/TISA difference were 13 to 817 microm/0.011 to 0.154 mm(2), with an average of 180 microm/0.073 mm(2). Multivariate regression analysis identified a positive correlation between anterior chamber depth and the AOD/TISA difference. Fifty eyes showed significant correlations between AOD/TISA and pupil diameter, whereas one eye showed no association. Four eyes in the narrow angle group developed appositional angle closure in the dark.
Conclusions: The dynamic dark-light changes of the anterior chamber angle can be imaged and analyzed with anterior segment OCT. Although the angle width generally decreased linearly with increasing pupil diameter, the differences of the angle width measured in the dark and in the light varied substantially among individuals.
Similar articles
-
Dynamic analysis of iris configuration with anterior segment optical coherence tomography.Invest Ophthalmol Vis Sci. 2010 Aug;51(8):4040-6. doi: 10.1167/iovs.09-3941. Epub 2010 Mar 17. Invest Ophthalmol Vis Sci. 2010. PMID: 20237248
-
Anterior chamber angle measurement with anterior segment optical coherence tomography: a comparison between slit lamp OCT and Visante OCT.Invest Ophthalmol Vis Sci. 2008 Aug;49(8):3469-74. doi: 10.1167/iovs.07-1477. Epub 2008 Apr 11. Invest Ophthalmol Vis Sci. 2008. PMID: 18408181
-
Evaluation of anterior chamber angle under dark and light conditions in angle closure glaucoma: An anterior segment OCT study.Cont Lens Anterior Eye. 2014 Aug;37(4):300-4. doi: 10.1016/j.clae.2014.04.002. Epub 2014 Jun 6. Cont Lens Anterior Eye. 2014. PMID: 24909857
-
Optical coherence tomography imaging of the anterior chamber angle.Ophthalmol Clin North Am. 2005 Sep;18(3):375-81, vi. doi: 10.1016/j.ohc.2005.05.007. Ophthalmol Clin North Am. 2005. PMID: 16054995 Review.
-
Anterior chamber angle assessment techniques.Surv Ophthalmol. 2008 May-Jun;53(3):250-73. doi: 10.1016/j.survophthal.2007.10.012. Surv Ophthalmol. 2008. PMID: 18501270 Review.
Cited by
-
Identification of Peripheral Anterior Synechia by Corneal Deformation Using Air-Puff Dynamic Anterior Segment Optical Coherence Tomography.Front Bioeng Biotechnol. 2022 Apr 1;10:856531. doi: 10.3389/fbioe.2022.856531. eCollection 2022. Front Bioeng Biotechnol. 2022. PMID: 35433648 Free PMC article.
-
Optical Coherence Tomography as a Tool for Ocular Dynamics Estimation.Biomed Res Int. 2015;2015:293693. doi: 10.1155/2015/293693. Epub 2015 Oct 18. Biomed Res Int. 2015. PMID: 26557659 Free PMC article.
-
Increase in anterior chamber angle depth after topical pilocarpine measured by spectral domain optical coherence tomography: A possible additional indicator for laser peripheral iridotomy in primary angle-closure suspects in an opportunistic set-up.Indian J Ophthalmol. 2022 Dec;70(12):4174-4179. doi: 10.4103/ijo.IJO_764_22. Indian J Ophthalmol. 2022. PMID: 36453309 Free PMC article. Clinical Trial.
-
Comparison of deep anterior lamellar keratoplasty and penetrating keratoplasty with respect to postoperative corneal sensitivity and tear film function.Graefes Arch Clin Exp Ophthalmol. 2014 Nov;252(11):1779-87. doi: 10.1007/s00417-014-2748-6. Epub 2014 Jul 31. Graefes Arch Clin Exp Ophthalmol. 2014. PMID: 25078353
-
Reproducibility and Agreement of Anterior Segment Parameter Measurements Obtained Using the CASIA2 and Spectralis OCT2 Optical Coherence Tomography Devices.J Glaucoma. 2017 Nov;26(11):974-979. doi: 10.1097/IJG.0000000000000788. J Glaucoma. 2017. PMID: 28930883 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources