Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study
- PMID: 17724263
- DOI: 10.1161/CIRCULATIONAHA.106.678664
Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study
Abstract
Background: The Prospective Evaluation of Pacemaker Lead Endocarditis study is a multicenter, prospective survey of the incidence and risk factors of infectious complications after implantation of pacemakers and cardioverter-defibrillators.
Methods and results: Between January 1, 2000, and December 31, 2000, 6319 consecutive recipients of implantable systems were enrolled at 44 medical centers and followed up for 12 months. All infectious complications were recorded, and their occurrence was related to the baseline demographic, clinical, and procedural characteristics. Among 5866 pacing systems, 3789 included 2 and 117 had >2 leads; among 453 implantable cardioverter-defibrillators, 178 were dual-lead systems. A total of 4461 de novo implantations occurred and 1858 pulse generator or lead replacements. Reinterventions were performed before hospital discharge in 101 patients. Single- and multiple-variable logistic regression analyses were performed to identify risk factors; adjusted odds ratios (aORs) and 95% confidence intervals (CIs) were calculated. At 12 months, device-related infections were reported in 42 patients (0.68%; 95% CI, 0.47 to 0.89). The occurrence of infection was positively correlated with fever within 24 hours before the implantation procedure (aOR, 5.83; 95% CI, 2.00 to 16.98), use of temporary pacing before the implantation procedure (aOR, 2.46; 95% CI, 1.09 to 5.13), and early reinterventions (aOR, 15.04; 95% CI, 6.7 to 33.73). Implantation of a new system (aOR, 0.46; 95% CI, 0.24 to 0.87) and antibiotic prophylaxis (aOR, 0.4; 95% CI, 0.18 to 0.86) were negatively correlated with risk of infection.
Conclusions: This study identified several factors of risk of device infection and confirmed the efficacy of antibiotic prophylaxis in recipients of new or replacement pacemakers or implantable cardioverter-defibrillators.
Similar articles
-
Efficacy of antibiotic prophylaxis before the implantation of pacemakers and cardioverter-defibrillators: results of a large, prospective, randomized, double-blinded, placebo-controlled trial.Circ Arrhythm Electrophysiol. 2009 Feb;2(1):29-34. doi: 10.1161/CIRCEP.108.795906. Epub 2009 Feb 10. Circ Arrhythm Electrophysiol. 2009. PMID: 19808441 Clinical Trial.
-
Device-related infection among patients with pacemakers and implantable defibrillators: incidence, risk factors, and consequences.J Cardiovasc Electrophysiol. 2010 Jul;21(7):786-90. doi: 10.1111/j.1540-8167.2009.01690.x. Epub 2010 Jan 22. J Cardiovasc Electrophysiol. 2010. PMID: 20102431
-
Permanent pacemaker and implantable cardioverter defibrillator infection: a population-based study.Arch Intern Med. 2007 Apr 9;167(7):669-75. doi: 10.1001/archinte.167.7.669. Arch Intern Med. 2007. PMID: 17420425
-
[Infectious complications of cardioverter/defibrillators--case report and review of the literature].Acta Med Austriaca. 1996;23(5):164-7. Acta Med Austriaca. 1996. PMID: 9082746 Review. German.
-
[Pacemaker and implantable cardioverter defibrillator infections].Med Mal Infect. 2010 Aug;40(8):429-39. doi: 10.1016/j.medmal.2009.11.005. Epub 2009 Dec 30. Med Mal Infect. 2010. PMID: 20044226 Review. French.
Cited by
-
Current trends in the management of cardiac implantable electronic device (CIED) infections.Intern Emerg Med. 2013 Sep;8(6):465-76. doi: 10.1007/s11739-012-0797-6. Epub 2012 Jun 29. Intern Emerg Med. 2013. PMID: 22744804 Review.
-
An unexpected cause of a swollen pacemaker pocket: a case report.Eur Heart J Case Rep. 2022 May 24;6(6):ytac211. doi: 10.1093/ehjcr/ytac211. eCollection 2022 Jun. Eur Heart J Case Rep. 2022. PMID: 35685031 Free PMC article.
-
Cardiac implantable electronic device infections: facts, current practice, and the unanswered questions.Curr Infect Dis Rep. 2014 Sep;16(9):425. doi: 10.1007/s11908-014-0425-x. Curr Infect Dis Rep. 2014. PMID: 25053116
-
Cardiac device-associated lead infection: a diagnosis not to be missed.Oxf Med Case Reports. 2015 Mar 6;2015(3):222-5. doi: 10.1093/omcr/omv014. eCollection 2015 Mar. Oxf Med Case Reports. 2015. PMID: 26421154 Free PMC article.
-
Risk of Cardiac Implantable Electronic Device Infection after Early versus Delayed Lead Repositioning.J Cardiovasc Dev Dis. 2024 Apr 9;11(4):117. doi: 10.3390/jcdd11040117. J Cardiovasc Dev Dis. 2024. PMID: 38667735 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical