Bioelectrical impedance: a new method for measuring post-traumatic swelling
- PMID: 17762477
- DOI: 10.1097/BOT.0b013e318123e88a
Bioelectrical impedance: a new method for measuring post-traumatic swelling
Abstract
Objectives: The aim of this pilot study was to investigate whether measurement of the bioelectrical impedance of the lower limb could be used to measure the swelling resulting from acute ankle fracture.
Methods: The impedance of each ankle was measured in 14 patients with isolated acute ankle fracture. The degree of ankle swelling was also directly assessed by measurement of the ankle circumference and diameter and by the water displacement method. A control group of 17 healthy subjects with uninjured ankles was similarly assessed.
Results: The impedance of the ankle was significantly reduced in patients with ankle fracture, and there was a strong inverse relationship between the degree of this reduction and the amount of swelling as directly measured. The relationship was stronger using the impedance method than the circumference and diameter methods.
Conclusions: We conclude that bioelectrical impedance can be used to measure ankle swelling in the presence of injury and could potentially be used both to monitor swelling clinically and as a research tool in studies of swelling management. More research is required to further define the potential role for this technique.
Similar articles
-
Reliability and minimal detectable change for the figure-of-eight-20 method of, measurement of ankle edema.J Orthop Sports Phys Ther. 2007 Apr;37(4):199-205. doi: 10.2519/jospt.2007.2371. J Orthop Sports Phys Ther. 2007. PMID: 17469673
-
Relationship between ankle invertor H-reflexes and acute swelling induced by inversion ankle sprain.J Orthop Sports Phys Ther. 1999 Jun;29(6):339-44. doi: 10.2519/jospt.1999.29.6.339. J Orthop Sports Phys Ther. 1999. PMID: 10370917
-
Comparison of acute swelling and function in subjects with lateral ankle injury.J Orthop Sports Phys Ther. 2001 Jul;31(7):384-8. doi: 10.2519/jospt.2001.31.7.384. J Orthop Sports Phys Ther. 2001. PMID: 11451309
-
Body fluid volumes measurements by impedance: A review of bioimpedance spectroscopy (BIS) and bioimpedance analysis (BIA) methods.Med Eng Phys. 2008 Dec;30(10):1257-69. doi: 10.1016/j.medengphy.2008.06.009. Epub 2008 Aug 3. Med Eng Phys. 2008. PMID: 18676172 Review.
-
Understanding lower leg volume measurements used in clinical studies focused on venous leg edema.Int Angiol. 2018 Dec;37(6):437-443. doi: 10.23736/S0392-9590.18.04057-9. Epub 2018 Sep 24. Int Angiol. 2018. PMID: 30256052 Review.
Cited by
-
Bioimpedance spectroscopy for swelling evaluation following total knee arthroplasty: a validation study.BMC Musculoskelet Disord. 2015 Apr 25;16:100. doi: 10.1186/s12891-015-0559-5. BMC Musculoskelet Disord. 2015. PMID: 25907994 Free PMC article.
-
Mid-Activity and At-Home Wearable Bioimpedance Elucidates an Interpretable Digital Biomarker of Muscle Fatigue.IEEE Trans Biomed Eng. 2023 Dec;70(12):3513-3524. doi: 10.1109/TBME.2023.3290530. Epub 2023 Nov 21. IEEE Trans Biomed Eng. 2023. PMID: 37405890 Free PMC article.
-
Difference in Bioimpedance Across the Knee in Un-Injured Young Adults.Iowa Orthop J. 2023;43(1):123-129. Iowa Orthop J. 2023. PMID: 37383845 Free PMC article.
-
Robust Method for Mid-Activity Tracking and Evaluation of Ankle Health Post-Injury.IEEE Trans Biomed Eng. 2021 Apr;68(4):1341-1350. doi: 10.1109/TBME.2020.3027477. Epub 2021 Mar 18. IEEE Trans Biomed Eng. 2021. PMID: 32997618 Free PMC article.