International Federation of Clinical Chemistry (IFCC) scientific division IFCC recommendation. Recommendation on sampling, transport and storage for the determination of the concentration of ionized calcium in whole blood, plasma and serum
- PMID: 1776715
International Federation of Clinical Chemistry (IFCC) scientific division IFCC recommendation. Recommendation on sampling, transport and storage for the determination of the concentration of ionized calcium in whole blood, plasma and serum
Abstract
The substance concentration of ionized calcium (cCa2+) in blood, plasma or serum preanalytically may be affected by pH changes of the sample, calcium binding by heparin, and dilution by the anticoagulant solution. pH changes in whole blood can be minimized by anaerobic sampling to avoid loss of CO2, by measuring as soon as possible or by storing the sample in iced water to avoid lactic acid formation. cCa2+ and pH should be determined simultaneously. Plasma or serum: if centrifuged in a closed tube and measured immediately the pH of the sample will be close to the original value. If delay has occurred between centrifugation and the measurement, causing substantial loss of CO2, equilibration of the sample with a gas mixture corresponding to PCO2 = 5.3 kPa prior to the measurement is recommended. Conversion of the measured values to cCa2+ (7.4) is only valid if the pH is in the range 7.2-7.6 Ca2+ binding by heparin can be minimized by using either of the following: a final concentration of sodium or lithium heparinate of 15 IU/ml blood or less, by use of calcium titrated heparin with a final concentration less than 50 IU/ml blood. Dilution effect can be avoided by use of dry heparin in capillaries or syringes. When heparin solutions are used errors due to dilution or calcium binding can be reduced using syringes with a heparin solution containing free calcium ions corresponding to the mean concentration of ionized calcium in normal plasma. Conditions for blood collection, storage, and transport to avoid preanalytical errors are described.
Similar articles
-
International Federation of Clinical Chemistry (IFCC), scientific division: IFCC recommendation on sampling transport and storage for the determination of the concentration of ionized calcium in whole blood, plasma and serum.Eur J Clin Chem Clin Biochem. 1991 Nov;29(11):767-72. Eur J Clin Chem Clin Biochem. 1991. PMID: 1782285
-
Recommendation on sampling, transport, and storage for the determination of the concentration of ionized calcium in whole blood, plasma, and serum. IFC Scientific Division, Working Group on Ion-Selective Electrodes (WGSE).J Int Fed Clin Chem. 1992 Sep;4(4):147-52. J Int Fed Clin Chem. 1992. PMID: 10148305
-
IFCC recommendation on sampling, transport and storage for the determination of the concentration of ionized calcium in whole blood, plasma and serum.J Automat Chem. 1991;13(5):235-9. doi: 10.1155/S1463924691000391. J Automat Chem. 1991. PMID: 18924910 Free PMC article.
-
[Determination of electrolytes in serum and plasma].Wien Klin Wochenschr Suppl. 1992;192:37-41. Wien Klin Wochenschr Suppl. 1992. PMID: 1502823 Review. German.
-
[Direct determination or estimated value of plasma ionized calcium : indications and limits].Ann Biol Clin (Paris). 2003 Jul-Aug;61(4):393-9. Ann Biol Clin (Paris). 2003. PMID: 12915348 Review. French.
Cited by
-
Influence of different storage times and temperatures on blood gas and acid-base balance in ovine venous blood.Open Vet J. 2013;3(1):1-7. Epub 2013 Jan 3. Open Vet J. 2013. PMID: 26623304 Free PMC article.
-
Ionized calcium measurement in serum and plasma by ion selective electrodes: comparison of measured and calculated parameters.Indian J Clin Biochem. 2014 Jul;29(3):327-32. doi: 10.1007/s12291-013-0360-x. Epub 2013 Jul 25. Indian J Clin Biochem. 2014. PMID: 24966481 Free PMC article.
-
A New Predictive Equation for Estimating Serum Ionized Calcium Levels in Patients on Chronic Hemodialysis.Med Sci Monit. 2023 Oct 9;29:e941321. doi: 10.12659/MSM.941321. Med Sci Monit. 2023. PMID: 37807497 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources