Percutaneous transtracheal emergency ventilation with a self-made device in an animal model
- PMID: 17767634
- DOI: 10.1111/j.1460-9592.2007.02244.x
Percutaneous transtracheal emergency ventilation with a self-made device in an animal model
Abstract
Background: Special equipment for emergency percutaneous transtracheal ventilation is often not immediately available. We used a self-made device consisting of a three-way stopcock connected between a G-15 transtracheal airway catheter and an oxygen supply in a simulated 'cannot intubate, cannot ventilate' scenario and tested the hypothesis that the effectiveness of the device depends on the body weight of the experimental animals.
Methods: With approval of the local animal protection committee, two groups of six pigs each with a body weight of 21 +/- 2 and 36 +/- 6 kg, respectively, were tracheally intubated and mechanically ventilated after induction of anesthesia. Hemodynamic monitoring included cardiac output and arterial, central venous and pulmonary artery pressures. An emergency transtracheal airway catheter was inserted into the trachea and a situation of partial expiratory airway obstruction was created. Each animal was ventilated for 15 min via the transtracheal airway catheter with the self-made device (FiO(2) 1.0 at an oxygen flow of 15 l x min(-1); respiratory rate of 60 min(-1); I/E ratio approximately 1:1).
Results: Whereas satisfactory oxygenation was achieved in all animals, sufficient ventilation was obtained only in the lightweight animals.
Conclusions: Adequate oxygenation could be provided by jet-like transtracheal ventilation with a simple self-made device using easily available materials. Sufficient ventilation depends on the body weight of the animals.
Comment in
-
Limitations of self-made jet devices.Paediatr Anaesth. 2008 Oct;18(10):984. doi: 10.1111/j.1460-9592.2008.02656.x. Paediatr Anaesth. 2008. PMID: 18811838 No abstract available.
Similar articles
-
Percutaneous transtracheal emergency ventilation during respiratory arrest: comparison of the oxygen flow modulator with a hand-triggered emergency jet injector in an animal model.Am J Emerg Med. 2006 Jul;24(4):455-9. doi: 10.1016/j.ajem.2006.01.014. Am J Emerg Med. 2006. PMID: 16787805
-
Successful transtracheal lung ventilation using a manual respiration valve: an in vitro and in vivo study.Anesthesiology. 2008 Aug;109(2):251-9. doi: 10.1097/ALN.0b013e31817f3de6. Anesthesiology. 2008. PMID: 18648234
-
[Preclinical transtracheal emergency ventilation. Animal experimental comparison of two techniques].Anaesthesist. 2007 Nov;56(11):1115-9. doi: 10.1007/s00101-007-1233-6. Anaesthesist. 2007. PMID: 17639282 German.
-
The critical airway, rescue ventilation, and the combitube: Part 2.AANA J. 2004 Apr;72(2):115-24. AANA J. 2004. PMID: 15098524 Review.
-
The critical airway, rescue ventilation, and the combitube: Part 1.AANA J. 2004 Feb;72(1):17-27. AANA J. 2004. PMID: 15098515 Review.
Cited by
-
Tidal volume and airway pressure under percutaneous transtracheal ventilation without a jet ventilator: comparison of high-flow oxygen ventilation and manual ventilation in complete and incomplete upper airway obstruction models.J Anesth. 2014 Jun;28(3):341-6. doi: 10.1007/s00540-013-1733-2. Epub 2013 Nov 9. J Anesth. 2014. PMID: 24212332
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous