Locking compression plate fixation of Vancouver type-B1 periprosthetic femoral fractures
- PMID: 17768193
- DOI: 10.2106/JBJS.F.01224
Locking compression plate fixation of Vancouver type-B1 periprosthetic femoral fractures
Abstract
Background: Fractures occurring at or near the distal tip of a hip prosthesis with a stable femoral stem (Vancouver type-B fractures) are associated with many complications because of the inherently unstable fracture pattern. Locking compression plates use screws that lock into the plate allowing multiple points of unicortical fixation. Such unicortical fixation may lower the risk of damage to the cement mantle or a stable femoral stem during the treatment of a periprosthetic femoral fracture. The purpose of this study was to analyze clinically and radiographically a group of patients with a Vancouver type-B1 periprosthetic femoral fracture treated with open reduction and internal fixation with use of a locking compression plate.
Methods: Fourteen consecutive patients (fourteen hips) with a Vancouver type-B1 periprosthetic femoral fracture were treated with a locking compression plate. There were five men and nine women with an average age of sixty-eight years at the time of fracture. All of the fractures occurred after a total hip arthroplasty performed with cement, and eleven of the arthroplasties were revisions. In addition to the plate, cortical strut allografts were used to stabilize five fractures. The patients were assessed clinically and radiographically.
Results: The average duration of follow-up was twenty months. Eight fractures healed uneventfully at an average of 5.4 months. Three treatment constructs failed with fracture of the plate within twelve months after surgery. An additional three constructs also failed because of plate pullout. All failures except one occurred in constructs in which a cortical strut allograft had not been utilized.
Conclusions: On the basis of the high failure rate in this series of patients, locking compression plates do not appear to offer advantages over other types of plates in the treatment of type-B1 periprosthetic femoral fractures. Despite the potential to preserve the cement mantle, the locked screws did not appear to offer good pullout resistance in this fracture type. We believe that supplementation with strut allografts should be used routinely if this type of locking compression plate is selected to treat these fractures.
Similar articles
-
High union rates of locking compression plating with cortical strut allograft for type B1 periprosthetic femoral fractures.Int Orthop. 2016 Nov;40(11):2365-2371. doi: 10.1007/s00264-015-3107-x. Epub 2016 Jan 13. Int Orthop. 2016. PMID: 26759257
-
Use of Locking Plate and Strut Onlay Allografts for Periprosthetic Fracture Around Well-Fixed Femoral Components.J Arthroplasty. 2017 Jan;32(1):166-170. doi: 10.1016/j.arth.2016.05.064. Epub 2016 Jun 7. J Arthroplasty. 2017. PMID: 27407037
-
Superior outcome of strut allograft-augmented plate fixation for the treatment of periprosthetic fractures around a stable femoral stem.Injury. 2013 Nov;44(11):1556-60. doi: 10.1016/j.injury.2013.04.025. Epub 2013 Jun 10. Injury. 2013. PMID: 23755973
-
Locking plate fixation for Vancouver B1 periprosthetic femoral fractures: a critical analysis of 135 cases.J Orthop Sci. 2013 May;18(3):426-36. doi: 10.1007/s00776-013-0359-4. Epub 2013 Feb 19. J Orthop Sci. 2013. PMID: 23420340 Review.
-
A retrospective review of the Dall-Miles plate for periprosthetic femoral fractures: twenty-seven cases and a review of the literature.Injury. 2014 Dec;45(12):1958-63. doi: 10.1016/j.injury.2014.08.034. Epub 2014 Aug 27. Injury. 2014. PMID: 25225174 Review.
Cited by
-
Treatment of periprosthetic femoral fractures after femoral revision using a long stem.BMC Musculoskelet Disord. 2015 May 10;16:113. doi: 10.1186/s12891-015-0565-7. BMC Musculoskelet Disord. 2015. PMID: 25958328 Free PMC article.
-
A review of periprosthetic femoral fractures associated with total hip arthroplasty.Geriatr Orthop Surg Rehabil. 2012 Sep;3(3):107-20. doi: 10.1177/2151458512462870. Geriatr Orthop Surg Rehabil. 2012. PMID: 23569704 Free PMC article.
-
Postoperative periprosthetic fractures in patients with an Exeter stem due to a femoral neck fracture: cumulative incidence and surgical outcome.Int Orthop. 2015 Sep;39(9):1683-8. doi: 10.1007/s00264-014-2570-0. Epub 2014 Oct 24. Int Orthop. 2015. PMID: 25341951
-
Interprosthetic femoral fractures treated with locking plate.Int Orthop. 2014 Oct;38(10):2183-9. doi: 10.1007/s00264-014-2414-y. Epub 2014 Jul 12. Int Orthop. 2014. PMID: 25011409
-
High Failure Rates of Locking Compression Plate Osteosynthesis with Transverse Fracture around a Well-Fixed Stem Tip for Periprosthetic Femoral Fracture.J Clin Med. 2020 Nov 22;9(11):3758. doi: 10.3390/jcm9113758. J Clin Med. 2020. PMID: 33266369 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical