The Vicenza wearable artificial kidney for peritoneal dialysis (ViWAK PD)
- PMID: 17785968
- DOI: 10.1159/000107775
The Vicenza wearable artificial kidney for peritoneal dialysis (ViWAK PD)
Abstract
Background: The study describes the structure and operational characteristics of a new wearable system for continuous ambulatory peritoneal dialysis (CAPD) for chronic kidney disease patients.
Methods: We designed a wearable system consisting of: (1) a double lumen peritoneal catheter; (2) a dialysate outflow line; (3) a miniaturized rotary pump; (4) a circuit for dialysate regeneration featuring a waterproof container with 4 cartridges in parallel with a mixture of activated carbon and polystyrenic resins; (5) a filter for deaeration and microbiological safety; (6) a dialysate inflow line, and (7) a handheld computer as a remote control. The system has been tested circulating 12 liters of exhausted PD solution through the experimental adsorption unit at a rate of 20 ml/min. Creatinine, beta2-microglobulin (beta2-MG) and angiogenin were measured before and after the adsorption unit at baseline, and after 4 and 10 h of use.
Results: The cartridges containing polystyrenic resin completely removed beta2-MG and angiogenin from the fluid batch. Those with the activated carbon removed completely urea and creatinine. The final result was 11.2 liters of net solute clearance. The system is designed to be used as follows: The peritoneal cavity is loaded in the morning with 2 liters of fresh PD solution. After 2 h, when dialysate/plasma equilibration at approximately 50% has occurred, recirculation is activated for 10 h at a rate of 20 ml/min. After this period, recirculation stops and glucose is optionally added to the peritoneal cavity to achieve ultrafiltration if needed. After 2 h the fluid is drained and a 2-liter icodextrin exchange is performed overnight to achieve further ultrafiltration. The clearance provided by the minicycler is further increased by the 2-liter exchange and the overnight exchange. Therefore, the system operates 24 h/day and provides creatinine and beta2-MG clearance in the range of 15-16 liters/day, corresponding to a weekly clearance of 100-110 liters. The patient reduces the number of exchanges compared to CAPD and uses less fluid than in automated peritoneal dialysis (APD). Furthermore, the handheld computer allows for prescription and assessment of the therapy providing information on cartridge saturation, flow and pressure conditions and offering the possibility of remote wireless control of operations. Some problems still remain to be solved in the present configuration including the addition of an injection system for glucose and bicarbonate when needed, a system to reduce fibrin delivery to the sorbent and finally a more complex mixture of sorbents to make sure a complete removal of small molecules including urea is achieved.
Conclusion: The wearable PD system may become a possible alternative to APD or CAPD reducing the time dedicated to perform exchanges and improving peritoneal dialysis adequacy and patient's rehabilitation.
Comment in
-
The long road to wearable blood-cleansing devices.Blood Purif. 2007;25(4):377-82. doi: 10.1159/000107774. Epub 2007 Sep 3. Blood Purif. 2007. PMID: 17785967 No abstract available.
Similar articles
-
Peritoneal membrane characteristics in patients on peritoneal dialysis.Saudi J Kidney Dis Transpl. 2011 Jan;22(1):49-53. Saudi J Kidney Dis Transpl. 2011. PMID: 21196612
-
The long road to wearable blood-cleansing devices.Blood Purif. 2007;25(4):377-82. doi: 10.1159/000107774. Epub 2007 Sep 3. Blood Purif. 2007. PMID: 17785967 No abstract available.
-
Recirculating peritoneal dialysis system using urease-fixed silk fibroin membrane filter with spherical carbonaceous adsorbent.Mater Sci Eng C Mater Biol Appl. 2019 Apr;97:55-66. doi: 10.1016/j.msec.2018.12.021. Epub 2018 Dec 8. Mater Sci Eng C Mater Biol Appl. 2019. PMID: 30678941
-
Wearable artificial kidneys for continuous dialysis. A personal view.ASAIO J. 1993 Jan-Mar;39(1):19-23. ASAIO J. 1993. PMID: 8439675 Review.
-
The future of the artificial kidney: moving towards wearable and miniaturized devices.Nefrologia. 2011;31(1):9-16. doi: 10.3265/Nefrologia.pre2010.Nov.10758. Nefrologia. 2011. PMID: 21270908 Review.
Cited by
-
Removal of urea by electro-oxidation in a miniature dialysis device: a study in awake goats.Am J Physiol Renal Physiol. 2018 Nov 1;315(5):F1385-F1397. doi: 10.1152/ajprenal.00094.2018. Epub 2018 Jul 11. Am J Physiol Renal Physiol. 2018. PMID: 29993277 Free PMC article.
-
Artificial kidney: Challenges and opportunities.World J Transplant. 2024 Mar 18;14(1):89025. doi: 10.5500/wjt.v14.i1.89025. World J Transplant. 2024. PMID: 38576754 Free PMC article. Review.
-
Portable and wearable dialysis devices for the treatment of patients with end-stage kidney failure: Wishful thinking or just over the horizon?Pediatr Nephrol. 2015 Dec;30(12):2053-60. doi: 10.1007/s00467-014-2968-3. Epub 2014 Oct 21. Pediatr Nephrol. 2015. PMID: 25330876 Free PMC article. Review.
-
The future of the artificial kidney.Indian J Urol. 2021 Oct-Dec;37(4):310-317. doi: 10.4103/iju.IJU_273_21. Epub 2021 Oct 1. Indian J Urol. 2021. PMID: 34759521 Free PMC article. Review.
-
Wearables in Nephrology: Fanciful Gadgetry or Prêt-à-Porter?Sensors (Basel). 2023 Jan 26;23(3):1361. doi: 10.3390/s23031361. Sensors (Basel). 2023. PMID: 36772401 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources