. 2008 Jan;22(1):214-20.
doi: 10.1007/s00464-007-9565-5. Epub 2007 Sep 3.

Complete endoscopic closure of gastrotomy after natural orifice translumenal endoscopic surgery using the NDO Plicator

Affiliations

Complete endoscopic closure of gastrotomy after natural orifice translumenal endoscopic surgery using the NDO Plicator

M F McGee et al. Surg Endosc. 2008 Jan.

Abstract

Background: The NDO Plicator is a device developed for endoscopic treatment of gastroesophageal reflux disease (GERD) by approximation of tissues together with a double-pledgeted U-stitch. It was theorized that this device may facilitate transgastric natural orifice translumenal endoscopic surgery (NOTES) because closure of the transgastric defect remains a key component for advancement of this new technology.

Methods: A standardized 12-mm gastrotomy was created endoscopically in four pigs using a combination of needle-knife cautery and balloon dilation. As the endoscope was removed, a Savary soft-tipped wire was introduced into the stomach, and the NDO Plicator was subsequently advanced over the wire. Each defect was identified, and the device was positioned. If necessary, the Plicator's tissue grasper was used to hold the superior aspect of the gastrotomy and bring the opposed borders of the defect within the jaws of the device. The device was fired three times, leaving three pledgeted suture bundles to close the gastric defect. After closure, each animal was explored, and the integrity of the closure was assessed. The animals underwent in vivo contrast fluoroscopy and ex vivo burst pressure testing studies for assessment of leakage at the closure site.

Results: The first animal was used to test feasibility, refine techniques, and develop a standard procedure. All of the next three animals studied showed complete sealing of the gastrotomy site without evidence of contrast extravasation on multiplanar fluoroscopic imaging. Each stomach was excised, submerged in water, and subjected to a pressurized air leak test. No leaks were noted until pressures exceeded 55 mmHg.

Conclusion: This study supports the use of the NDO Plicator for closure of standardized gastric defects in a porcine model. In addition to closing NOTES gastrotomies, the NDO Plicator may be a particularly useful tool for obtaining complete closure of gastric perforations and anastomotic leaks, and for performing stomal reduction after gastric bypass procedures. The mechanical properties of a closure are not the only factor determining whether a leak will develop. Tissue opposition, ischemia, and tension are important factors that are not easily or reliably measured. The physiologic relevance of gastric bursting pressure is not known. Therefore, corollary animal studies with longer-term evaluation are necessary before research proceeds to clinical trials.

PubMed Disclaimer

References

    1. Gastrointest Endosc. 2003 Nov;58(5):770-6 - PubMed
    1. Eur Surg Res. 2005 Mar-Apr;37(2):123-8 - PubMed
    1. Surg Endosc. 2006 Aug;20(8):1179-92 - PubMed
    1. Surg Endosc. 2006 Feb;20(2):329-33 - PubMed
    1. Am J Respir Crit Care Med. 2003 Sep 15;168(6):714-7 - PubMed