Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2007 Jul;12(4):233-41.
doi: 10.3109/10929080701552894.

A CT-free, intra-operative planning and navigation system for minimally invasive anterior spinal surgery - an accuracy study

Affiliations
Free article

A CT-free, intra-operative planning and navigation system for minimally invasive anterior spinal surgery - an accuracy study

B Maier et al. Comput Aided Surg. 2007 Jul.
Free article

Abstract

Objective: A comprehensive study was performed to evaluate the accuracy of a newly developed CT-free, intra-operative planning and navigation system for anterior spine surgery.

Materials and methods: Instruments and an image intensifier were tracked using the SurgiGATE navigation system. A laboratory study was performed on 27 plastic vertebrae. Fiducial markers were implanted in the vertebrae for accuracy evaluation purposes, and a dynamic reference base was placed on the vertebrae to establish a patient coordinate system (P-COS). Two fluoroscopic images were used for intra-operative planning. The graft bed plan was recorded in P-COS, followed by surgical formation of the graft bed, which was visualized. To evaluate the accuracy, the vertebrae were scanned with CT, and the markers were used to calculate an accurate paired-point registered transformation between the CT coordinate system and P-COS.

Results: Using the new SPO module, accurate planning and navigation of a resection of the vertebral body is possible using two fluoroscopic images. The overall mean error between the planned resection volume and the actual resection was 0.98 mm. In addition, the module can serve as an educational tool for training spine surgeons.

Conclusions: The new fluoroscopy-based system can be used safely for accurate performance of anterior resection during spondylodesis. New methods for safe and accurate registration during anterior spine surgery need to be developed.