Enhancement of subharmonic emission from encapsulated microbubbles by using a chirp excitation technique
- PMID: 17804880
- DOI: 10.1088/0031-9155/52/18/005
Enhancement of subharmonic emission from encapsulated microbubbles by using a chirp excitation technique
Abstract
Subharmonic contrast imaging promises to improve ultrasound-imaging quality by taking advantage of an increased contrast to tissue signal. However, acoustic pressures beyond the subharmonic generation threshold using common ultrasound pulses may induce significant contrast microbubble destruction. In this work, a chirp excitation technique is presented to enhance the subharmonic emission from encapsulated microbubbles. Chirp signals with a center frequency of 5 MHz, variable frequency range and duration time are employed to drive microbubbles in numerical simulation and experimental studies. We provide a theoretical evaluation of the chirp excitation pressure threshold and the acoustic pressure dependence of subharmonic based on Church's model and demonstrate that the amplitude and axial resolution of the subharmonic can be optimized by proper selection of the frequency range and time duration of the chirp signal. Measurements are qualitatively in agreement with the simulation. Moreover, we demonstrate that chirp excitation may be able to improve the amplitude of the subharmonic component up to 22 dB over the pulse excitation. The chirp excitation technique could potentially be used for improving the subharmonic contrast imaging quality.
Similar articles
-
A dual-frequency excitation technique for enhancing the sub-harmonic emission from encapsulated microbubbles.Phys Med Biol. 2009 Jul 7;54(13):4257-72. doi: 10.1088/0031-9155/54/13/019. Epub 2009 Jun 17. Phys Med Biol. 2009. PMID: 19531846
-
Contrast agent response to chirp reversal: simulations, optical observations, and acoustical verification.IEEE Trans Ultrason Ferroelectr Freq Control. 2009 Jun;56(6):1199-206. doi: 10.1109/TUFFC.2009.1161. IEEE Trans Ultrason Ferroelectr Freq Control. 2009. PMID: 19574127
-
Dual high-frequency difference excitation for contrast detection.IEEE Trans Ultrason Ferroelectr Freq Control. 2008 Oct;55(10):2164-76. doi: 10.1109/TUFFC.916. IEEE Trans Ultrason Ferroelectr Freq Control. 2008. PMID: 18986865
-
Broadband reduction of the second harmonic distortion during nonlinear ultrasound wave propagation.Ultrasound Med Biol. 2010 Oct;36(10):1568-80. doi: 10.1016/j.ultrasmedbio.2010.06.006. Ultrasound Med Biol. 2010. PMID: 20800962 Review.
-
Microbubble contrast for radiological imaging: 1. Principles.Ultrasound Q. 2006 Mar;22(1):5-13. Ultrasound Q. 2006. PMID: 16641788 Review.
Cited by
-
The delayed onset of subharmonic and ultraharmonic emissions from a phospholipid-shelled microbubble contrast agent.Ultrasound Med Biol. 2014 Apr;40(4):727-38. doi: 10.1016/j.ultrasmedbio.2014.01.002. Ultrasound Med Biol. 2014. PMID: 24582298 Free PMC article.
-
Effect of Pulse Shaping on Subharmonic Aided Pressure Estimation In Vitro and In Vivo.J Ultrasound Med. 2017 Jan;36(1):3-11. doi: 10.7863/ultra.15.11106. Epub 2016 Dec 10. J Ultrasound Med. 2017. PMID: 27943411 Free PMC article.
-
Combining Subharmonic and Ultraharmonic Modes for Intravascular Ultrasound Imaging: A Preliminary Evaluation.Ultrasound Med Biol. 2017 Nov;43(11):2725-2732. doi: 10.1016/j.ultrasmedbio.2017.07.012. Epub 2017 Aug 26. Ultrasound Med Biol. 2017. PMID: 28847499 Free PMC article.
-
Improving the sensitivity of high-frequency subharmonic imaging with coded excitation: a feasibility study.Med Phys. 2012 Apr;39(4):2049-60. doi: 10.1118/1.3694101. Med Phys. 2012. PMID: 22482626 Free PMC article.
-
Optimization of multi-pulse sequences for nonlinear contrast agent imaging using a cMUT array.Phys Med Biol. 2015 Apr 21;60(8):3111-27. doi: 10.1088/0031-9155/60/8/3111. Epub 2015 Mar 24. Phys Med Biol. 2015. PMID: 25803232 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources