Early experience in the use of quantitative image quality measurements for the quality assurance of full field digital mammography x-ray systems
- PMID: 17804881
- DOI: 10.1088/0031-9155/52/18/006
Early experience in the use of quantitative image quality measurements for the quality assurance of full field digital mammography x-ray systems
Abstract
Quantitative image quality results in the form of the modulation transfer function (MTF), normalized noise power spectrum (NNPS) and detective quantum efficiency (DQE) are presented for nine full field digital mammography (FFDM) systems. These parameters are routinely measured as part of the quality assurance (QA) programme for the seven FFDM units covered by our centre. Just one additional image is required compared to the standard FFDM protocol; this is the image of an edge, from which the MTF is calculated. A variance image is formed from one of the flood images used to measure the detector response and this provides useful information on the condition of the detector with respect to artefacts. Finally, the NNPS is calculated from the flood image acquired at a target detector air kerma (DAK) of 100 microGy. DQE is then estimated from these data; however, no correction is currently made for effects of detector cover transmission on DQE. The coefficient of variation (cov) of the 50% point of the MTF for five successive MTF results was 1%, while the cov for the 50% MTF point for an a-Se system over a period of 17 months was approximately 3%. For four a-Se based systems, the cov for the NNPS at 1 mm(-1) for a target DAK of 100 microGy was approximately 4%; the same result was found for four CsI based FFDM units. With regard to the stability of NNPS over time, the cov for four NNPS results acquired over a period of 12 months was also approximately 4%. The effect of acquisition geometry on NNPS was also assessed for a CsI based system. NNPS data acquired with the antiscatter grid in place showed increased noise at low spatial frequency; this effect was more severe as DAK increased. DQE results for the three detector types (a-Se, CsI and CR) are presented as a function of DAK. Some reduction in DQE was found for both the a-Se and CsI based systems at a target DAK of 12.5 microGy when compared to DQE data acquired at 100 microGy. For the CsI based systems, DQE at 1 mm(-1) fell from 0.49 at 100 microGy to 0.38 at 12.5 microGy. For the a-Se units, there was a slightly greater reduction in average DQE at 1 mm(-1), from 0.53 at 100 microGy to 0.31 at 12.5 microGy. Somewhat different behaviour was seen for the CR unit; DQE (at 1 mm(-1)) increased from 0.40 at 100 microGy to 0.49 at 12.5 microGy; however, DQE fell to 0.30 at 420 microGy. DQE stability over time was assessed using the cov of DQE at 1 mm(-1) and a target DAK of 100 microGy; the cov for data acquired over a period of 17 months for an a-Se system was approximately 7%. For comparison with conventional testing methods, the cov was calculated for contrast-detail (cd) data acquired over the same period of time for this unit. The cov for the threshold contrast results (averaged for disc diameters between 0.1 mm and 2 mm) was 6%, indicating similar stability.
Similar articles
-
A comparison between objective and subjective image quality measurements for a full field digital mammography system.Phys Med Biol. 2006 May 21;51(10):2441-63. doi: 10.1088/0031-9155/51/10/006. Epub 2006 Apr 26. Phys Med Biol. 2006. PMID: 16675862
-
Detective quantum efficiency measured as a function of energy for two full-field digital mammography systems.Phys Med Biol. 2009 May 7;54(9):2845-61. doi: 10.1088/0031-9155/54/9/017. Epub 2009 Apr 21. Phys Med Biol. 2009. PMID: 19384004
-
Retrospective analysis of a detector fault for a full field digital mammography system.Phys Med Biol. 2006 Nov 7;51(21):5655-73. doi: 10.1088/0031-9155/51/21/018. Epub 2006 Oct 16. Phys Med Biol. 2006. PMID: 17047276
-
Implementing digital quality control in a breast center.J Am Coll Radiol. 2004 Nov;1(11):854-60. doi: 10.1016/j.jacr.2004.05.006. J Am Coll Radiol. 2004. PMID: 17411717 Review.
-
Cassette-based digital mammography.Technol Cancer Res Treat. 2004 Oct;3(5):413-27. doi: 10.1177/153303460400300502. Technol Cancer Res Treat. 2004. PMID: 15453806 Review.
Cited by
-
A Novel Method to Model Image Creation Based on Mammographic Sensors Performance Parameters: A Theoretical Study.Sensors (Basel). 2023 Feb 20;23(4):2335. doi: 10.3390/s23042335. Sensors (Basel). 2023. PMID: 36850937 Free PMC article.
-
Imaging and Dosimetric Study on Direct Flat-Panel Detector-Based Digital Mammography System.J Med Phys. 2018 Oct-Dec;43(4):255-263. doi: 10.4103/jmp.JMP_64_18. J Med Phys. 2018. PMID: 30636851 Free PMC article.
-
Study of DQE dependence with beam quality on GE essential mammography flat panel.J Appl Clin Med Phys. 2010 Nov 25;12(1):3176. doi: 10.1120/jacmp.v12i1.3176. J Appl Clin Med Phys. 2010. PMID: 21330969 Free PMC article.
-
Investigation of noise sources for digital radiography systems.Radiol Phys Technol. 2017 Jun;10(2):171-179. doi: 10.1007/s12194-016-0381-2. Epub 2016 Oct 1. Radiol Phys Technol. 2017. PMID: 27696210
-
On the relevance of modulation transfer function measurements in digital mammography quality control.J Med Imaging (Bellingham). 2021 Mar;8(2):023505. doi: 10.1117/1.JMI.8.2.023505. Epub 2021 Apr 27. J Med Imaging (Bellingham). 2021. PMID: 33937435 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous