Factors affecting hip range of motion in surface replacement arthroplasty
- PMID: 17870221
- DOI: 10.1016/j.clinbiomech.2007.07.007
Factors affecting hip range of motion in surface replacement arthroplasty
Abstract
Background: Surface replacement arthroplasty aims to re-create normal hip biomechanics; however the pathoanathomy of the hip, prosthetic component design, surgical technique and other factors may have a significant impact on the range of motion restoration attained following surface replacement arthroplasty. However, there is paucity of information on the effect of such factors.
Methods: A computerized three-dimensional hip model was created from preoperative computerized tomography images of a patient who was scheduled for a surface replacement arthroplasty. The effects of the femoral component size, translation and orientation on the range of motion were analysed as was the effect of increasing the seating depth and modification of the version of the acetabular component.
Findings: Increasing the femoral component size led to global improvement in range of motion while translation increased range of motion in one direction but reduced it in the opposite direction. Change in the femoral component orientation had minimal effects on range of motion in comparison to the effect of changes in the version of the acetabular component. Increasing the seating depth of the acetabulum only caused reduced range of motion in internal rotation in 90 degrees flexion.
Interpretation: To restore hip range of motion, surgeons performing surface replacement arthroplasty should aim to reproduce the natural femoral head-neck offset. Although increasing the femoral component size may achieve this, more acetabular bone will be resected. Knowing the specific zones of impingement of each arc of movement, selective translation of the femoral component or femoral neck osteoplasty can restore femoral neck offset in more critical areas without affecting acetabular bone stock. Over deepening of the acetabulum or leaving rim osteophytes should also be avoided to prevent impingement.
Similar articles
-
Analysis of optimal range of socket orientations in total hip arthroplasty with use of computer-aided design simulation.J Orthop Res. 1998 Jul;16(4):513-7. doi: 10.1002/jor.1100160418. J Orthop Res. 1998. PMID: 9747795
-
Femoral head to neck offset after hip resurfacing is critical for range of motion.Clin Biomech (Bristol). 2012 Feb;27(2):165-9. doi: 10.1016/j.clinbiomech.2011.08.013. Epub 2011 Sep 16. Clin Biomech (Bristol). 2012. PMID: 21925779
-
The impact of the CCD-angle on range of motion and cup positioning in total hip arthroplasty.Clin Biomech (Bristol). 2005 Aug;20(7):723-8. doi: 10.1016/j.clinbiomech.2005.04.003. Clin Biomech (Bristol). 2005. PMID: 15964112
-
Metal-on-metal hip resurfacing: a skeptic's view.Clin Orthop Relat Res. 2007 Dec;465:86-91. doi: 10.1097/BLO.0b013e3181468911. Clin Orthop Relat Res. 2007. PMID: 17632416 Review.
-
Containment versus impingement: finding a compromise for cup placement in total hip arthroplasty.Int Orthop. 2007 Aug;31 Suppl 1(Suppl 1):S29-33. doi: 10.1007/s00264-007-0429-3. Int Orthop. 2007. PMID: 17661036 Free PMC article. Review.
Cited by
-
Digital templating and preoperative deformity analysis with standard imaging software.Clin Orthop Relat Res. 2009 Oct;467(10):2695-704. doi: 10.1007/s11999-009-0858-y. Epub 2009 May 15. Clin Orthop Relat Res. 2009. PMID: 19444527 Free PMC article.
-
Correlation between groin pain and cup design of hip-resurfacing implants: a prospective study.Int Orthop. 2014 May;38(5):923-8. doi: 10.1007/s00264-013-2226-5. Epub 2013 Dec 10. Int Orthop. 2014. PMID: 24323353 Free PMC article.
-
[Effect of Jumbo cups for total hip arthroplasty revision on femoral offset and hip center of rotation].Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2017 Feb 15;31(2):139-143. doi: 10.7507/1002-1892.201607089. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2017. PMID: 29786242 Free PMC article. Chinese.
-
Hip arthroscopy in patients with painful hip following resurfacing arthroplasty.Knee Surg Sports Traumatol Arthrosc. 2011 Sep;19(9):1514-20. doi: 10.1007/s00167-011-1463-7. Epub 2011 Mar 16. Knee Surg Sports Traumatol Arthrosc. 2011. PMID: 21409469
-
[Effects of femoral offset and mechanical axis of the lower extremity on hip after osteotomy for adult developmental dysplasia of the hip].Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2021 Dec 15;35(12):1549-1554. doi: 10.7507/1002-1892.202108080. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2021. PMID: 34913311 Free PMC article. Chinese.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical