Evaluation of the design and reliability of three elastomeric and one mechanical infusers
- PMID: 17873107
- DOI: 10.1177/1078155207078349
Evaluation of the design and reliability of three elastomeric and one mechanical infusers
Abstract
Background: Disposable elastomeric and mechanical (spring mechanism) infusers with pre-fixed flow rates are principally used for long-term antibiotic therapy, chemotherapy, antiviral therapy, and chronic pain management.
Methods: Three elastomeric pumps and one mechanical infuser pump are evaluated. The design and function are assessed using specific questionnaires elaborated for the pharmaceutical staff, nurses and patients, respectively. Several tests are performed to determine the flow accuracy and the influence of different parameters such as the temperature, the viscosity, the height, and the restrictor variability.
Results: The functional tests show that the perfusion duration is shorter than that specified by the manufacturers, the restrictors are not homogeneous within a batch. However, the temperature of the regulator, the viscosity of the solution, and the height of the reservoir influence the flow rate.
Conclusions: The ideal infuser should be small, light, unnoticeable, easy to fill, well identified and have a flow rate barely affected by external factors. None of the infusers has fulfilled all criteria, therefore the choice of infusers depends upon the nature of the treatment. The tested elastomeric infusers are adequate for 5-fluorouracil, but not for a treatment requiring a stable administration rate.
Similar articles
-
Elastomeric pump reliability in postoperative regional anesthesia: a survey of 430 consecutive devices.Anesth Analg. 2008 Dec;107(6):2079-84. doi: 10.1213/ane.0b013e318187c9bb. Anesth Analg. 2008. PMID: 19020162
-
[Current status of portable disposable infusers].Masui. 2006 Sep;55(9):1118-27. Masui. 2006. PMID: 16984010 Review. Japanese.
-
Effects of flow rate accuracy in two-day anticancer drug infusion with disposable pumps on plasma drug concentrations.Technol Health Care. 2024;32(3):1351-1360. doi: 10.3233/THC-230227. Technol Health Care. 2024. PMID: 37781825
-
Regional anaesthesia elastomeric pump performance after a single use and subsequent refill: a laboratory study.Anaesthesia. 2009 Jul;64(7):770-5. doi: 10.1111/j.1365-2044.2009.05941.x. Anaesthesia. 2009. PMID: 19624633
-
Disposable infusion pumps.Am J Health Syst Pharm. 2006 Jul 1;63(13):1260-8. doi: 10.2146/ajhp050408. Am J Health Syst Pharm. 2006. PMID: 16790578 Review.
Cited by
-
Mechanical performances of elastomers used in diffusers.Med Devices (Auckl). 2011;4:71-6. doi: 10.2147/MDER.S18633. Epub 2011 Jun 14. Med Devices (Auckl). 2011. PMID: 22915932 Free PMC article.
-
The flow Rate Accuracy of Elastomeric Infusion Pumps After Repeated Filling.Anesth Pain Med. 2014 Apr 7;4(2):e14989. doi: 10.5812/aapm.14989. eCollection 2014 May. Anesth Pain Med. 2014. PMID: 24829882 Free PMC article.
-
Efficacy and safety of a subacromial continuous ropivacaine infusion for post-operative pain management following arthroscopic rotator cuff surgery: a protocol for a randomised double-blind placebo-controlled trial.BMC Musculoskelet Disord. 2008 Apr 22;9:56. doi: 10.1186/1471-2474-9-56. BMC Musculoskelet Disord. 2008. PMID: 18430210 Free PMC article. Clinical Trial.
-
Acute profound abciximab induced thrombocytopenia: a correct management of a methodological error.BMJ Case Rep. 2009;2009:bcr12.2008.1381. doi: 10.1136/bcr.12.2008.1381. Epub 2009 Oct 5. BMJ Case Rep. 2009. PMID: 21977060 Free PMC article.
-
Protocol for a randomised crossover trial to evaluate patient and nurse satisfaction with electronic and elastomeric portable infusion pumps for the continuous administration of antibiotic therapy in the home: the Comparing Home Infusion Devices (CHID) study.BMJ Open. 2017 Jul 31;7(7):e016763. doi: 10.1136/bmjopen-2017-016763. BMJ Open. 2017. PMID: 28760798 Free PMC article. Clinical Trial.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources