Experimental validation of a versatile system of CT dosimetry using a conventional ion chamber: beyond CTDI100
- PMID: 17879802
- DOI: 10.1118/1.2757084
Experimental validation of a versatile system of CT dosimetry using a conventional ion chamber: beyond CTDI100
Abstract
This article is an experimental demonstration and authentication of a new method of computed tomography dosimetry [R. L. Dixon, Med. Phys. 30, 1272-1280 (2003)], which utilizes a short, conventional ion chamber rather than a pencil chamber, and which is more versatile than the latter. The value of CTDI100 correctly predicts the accumulated dose only for a total scan length L equal to 100 mm and underestimates the limiting equilibrium dose approached for longer, clinically relevant body scan lengths [R. L. Dixon, Med. Phys. 30, 1272-1280 (2003); K. D. Nakonechny, B. G. Fallone, and S. Rathee, Med. Phys. 32, 98-109 (2005); S. Mori, M. Endo, K. Nishizawa, T. Tsunoo, T. Aoyama, H. Fujiwara, and K. Murase, Med. Phys. 32, 1061-1069 (2005); R. L. Dixon, M. T. Munley, and E. Bayram, Med. Phys. 32, 3712-3728 (2005); R. L. Dixon, Med. Phys. 33, 3973-3976 (2006)]. Dixon [Med. Phys. 30, 1272-1280 (2003)] originally proposed an alternative using a short ion chamber and a helical scan acquisition to collect the same integral for any scan length L (and not limited 100 mm). The primary purpose of this work is to demonstrate experimentally the implementation, robustness, and versatility of this small ion chamber method in measuring the accumulated dose in the body phantom for any desired scan length L (up to the available phantom length) including the limiting equilibrium dose (symbolically CTDIinfinity), and validation of the method against the pencil chamber methodology. Additionally, a simple and robust method for independently verifying the active length of a pencil chamber is described. The results of measurements made in a 400 mm long, 32 cm diameter polymethylmethacrylate body phantom using a small Farmer-type ion chamber and two pencil chambers of lengths l=100 and 150 mm confirm that the two methodologies provide the same dose values at the corresponding scan lengths L=l. The measured equilibrium doses obtained for GE MDCT scanners at 120 kVp are CTDIinfinity = 1.75 CTDI100 on the central axis and 1.22 CTDI100 on the peripheral axes, illustrating a nontrivial shortfall of CTDI100 in that regard and in good agreement with comparable data [S. Mori, M. Endo, K. Nishizawa, T. Tsunoo, T. Aoyama, H. Fujiwara, and K. Murase, Med. Phys. 32, 1061-1069 (2005); J. M. Boone, Med. Phys. 34, 1364-1371 (2007)].
Similar articles
-
Cone beam CT dosimetry: a unified and self-consistent approach including all scan modalities--with or without phantom motion.Med Phys. 2010 Jun;37(6):2703-18. doi: 10.1118/1.3395578. Med Phys. 2010. PMID: 20632581 Free PMC article.
-
Novel methods of measuring single scan dose profiles and cumulative dose in CT.Med Phys. 2005 Jan;32(1):98-109. doi: 10.1118/1.1835571. Med Phys. 2005. PMID: 15719960
-
Calculations of two new dose metrics proposed by AAPM Task Group 111 using the measurements with standard CT dosimetry phantoms.Med Phys. 2013 Aug;40(8):081914. doi: 10.1118/1.4813899. Med Phys. 2013. PMID: 23927328
-
CT Dosimetry: What Has Been Achieved and What Remains to Be Done.Invest Radiol. 2021 Jan;56(1):62-68. doi: 10.1097/RLI.0000000000000727. Invest Radiol. 2021. PMID: 32932380 Review.
-
Dosimetry in Micro-computed Tomography: a Review of the Measurement Methods, Impacts, and Characterization of the Quantum GX Imaging System.Mol Imaging Biol. 2017 Aug;19(4):499-511. doi: 10.1007/s11307-016-1026-x. Mol Imaging Biol. 2017. PMID: 27957647 Free PMC article. Review.
Cited by
-
Computed tomography dosimetry with high-resolution detectors commonly used in radiotherapy - an energy dependence study.J Appl Clin Med Phys. 2015 Sep 8;16(5):396–407. doi: 10.1120/jacmp.v16i5.5302. J Appl Clin Med Phys. 2015. PMID: 26699294 Free PMC article.
-
Reducing radiation dose to selected organs by selecting the tube start angle in MDCT helical scans: a Monte Carlo based study.Med Phys. 2009 Dec;36(12):5654-64. doi: 10.1118/1.3259773. Med Phys. 2009. PMID: 20095278 Free PMC article.
-
Influence of difference in cross-sectional dose profile in a CTDI phantom on X-ray CT dose estimation: a Monte Carlo study.Radiol Phys Technol. 2014 Jan;7(1):133-40. doi: 10.1007/s12194-013-0242-1. Epub 2013 Nov 24. Radiol Phys Technol. 2014. PMID: 24272457
-
Assessment of radiation dose from abdominal quantitative CT with short scan length.Br J Radiol. 2017 Jun;90(1074):20160931. doi: 10.1259/bjr.20160931. Epub 2017 May 23. Br J Radiol. 2017. PMID: 28402125 Free PMC article.
-
In vitro dose measurements in a human cadaver with abdomen/pelvis CT scans.Med Phys. 2014 Sep;41(9):091911. doi: 10.1118/1.4893499. Med Phys. 2014. PMID: 25186398 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical