Analytic evaluation of the beta-human chorionic gonadotropin assay on the Abbott IMx and Elecsys2010 for its use in doping control
- PMID: 17884034
- DOI: 10.1016/j.clinbiochem.2007.06.015
Analytic evaluation of the beta-human chorionic gonadotropin assay on the Abbott IMx and Elecsys2010 for its use in doping control
Abstract
Objectives: The principal objective of this study was to compare the analytical performance of the Elecsys2010 (Roche Diagnostics) system with the IMx (Abbott laboratories) system for beta-hCG assay in order to assess its possible utility as a confirmation test for the quantitative measurement of beta-hCG in urine for doping control purposes.
Design and methods: Urine samples with spiked standard known concentrations of beta-hCG and different urine samples from athletes were used in order to determine the calibration curve stability and linearity, detection limit, total, within-run and between-run precision, and method comparison for the IMx and Elecsys2010 systems for beta-hCG assay, along with the stability of samples, at room temperature and at 4 degrees C.
Results: The IMx assay was linear up to 500 IU/L, whereas the Elecsys2010 assay was linear up to 1000 IU/L. The detection limit for the IMx and Elecsys2010 systems were 0.75 IU/L and 0.25 IU/L, respectively. The total precision of the IMx and Elecsys2010 systems were <or=5% for beta-hCG concentrations ranging from 20 to 900 IU/L and from 6.67 to 55.2 IU/L, respectively. The within-run precision tests for the IMx and Elecsys2010 systems yielded CV results of 2-3% and 3-4% respectively, while the between-run precision for the Elecsys2010 was less than 4%, whereas that of IMx was 9-16%. Regression analysis with spiked urine samples yielded the following: slope (SD) 0.7104 (0.003) and y-intercept (SD) -0.4286 (1.397) (Sy.x=2.328; r=0.9999; n=5), whereas regression analysis using the athletes' urine samples yielded the following results: slope (SD) 0.3692 (0.003) and y-intercept (SD) 0.0016 (0.003) (Sy.x=0.0096; r=0.9924; n=100). Sample stability studies evidenced a nonsignificant effect at 4 degrees C and room temperature (25 degrees C) for up to 1 month.
Conclusions: For doping control, the IMx and Elecys2010 may be used to screen and confirm the presence of beta-hCG in urine of athletes, respectively. Additional studies are needed to verify this finding. This study also underscores the need for beta-hCG assays standardization for doping control.
Similar articles
-
Evaluation of two immunoassays for the measurement of human chorionic gonadotropin in urine for anti-doping purposes.Clin Lab. 2010;56(5-6):197-206. Clin Lab. 2010. PMID: 20575467
-
Quantitation of whole molecule human chorionic gonadotropin and free beta-human chorionic gonadotropin subunit in the Abbott IMx analyser.Eur J Clin Chem Clin Biochem. 1991 Jan;29(1):77-80. Eur J Clin Chem Clin Biochem. 1991. PMID: 1710936
-
Assessment of the Abbott IMx assay system for the measurement of human chorionic gonadotropin levels in the treatment of ectopic pregnancy.Arch Pathol Lab Med. 1993 Jul;117(7):701-3. Arch Pathol Lab Med. 1993. PMID: 8323432
-
Immunoprocedures for detecting human chorionic gonadotropin: clinical aspects and doping control.Clin Chem. 1997 Jul;43(7):1293-8. Clin Chem. 1997. PMID: 9216477 Review.
-
A world apart: Inaccuracies of laboratory methodologies in antidoping testing.Clin Chim Acta. 2010 Aug 5;411(15-16):1003-8. doi: 10.1016/j.cca.2010.03.039. Epub 2010 Mar 31. Clin Chim Acta. 2010. PMID: 20362564 Review.
Cited by
-
Gonadotropins in doping: pharmacological basis and detection of illicit use.Br J Pharmacol. 2008 Jun;154(3):569-83. doi: 10.1038/bjp.2008.102. Epub 2008 Apr 14. Br J Pharmacol. 2008. PMID: 18414398 Free PMC article. Review.
-
Verification of quantitative analytical methods in medical laboratories.J Med Biochem. 2021 Jun 5;40(3):225-236. doi: 10.5937/jomb0-24764. J Med Biochem. 2021. PMID: 34177366 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources