Full-field digital versus screen-film mammography: comparative accuracy in concurrent screening cohorts
- PMID: 17885057
- DOI: 10.2214/AJR.07.2303
Full-field digital versus screen-film mammography: comparative accuracy in concurrent screening cohorts
Abstract
Objective: The purpose of this study was to compare the diagnostic accuracy of digital mammography with that of screen-film mammography in concurrent cohorts participating in the same population-based screening program.
Materials and methods: In a retrospective study covering 2004-2005, we compared digital with screen-film mammography in two concurrent screening cohorts of women 50-69 years old participating in a screening program operated from mobile units. Each cohort had 14,385 participants matched by age and interpreting radiologist from all participants consecutively registered. We compared recall and cancer detection rates.
Results: The recall rate was higher for digital mammography (4.56% vs 3.96%, p = 0.01), particularly when clustered microcalcifications were the only finding (1.05% vs 0.41%, p = 10(-6)) and for younger women (50-59 vs 60-69 years, 5.12% vs 4.17%, p = 0.009). The higher recall rate for digital mammography was mainly evident at incidence screening. The recall rate due to poor technical quality was lower with digital mammography (0.27% vs 0.50%, p = 0.002), possibly because real-time feedback was available. The detection rate was higher for digital mammography (0.72% vs 0.58%, p = 0.14), particularly for cancers depicted as clustered microcalcifications (0.26% vs 0.12%, p = 0.007), in younger (50-59 years) women (0.63% vs 0.42%, p = 0.09), and in denser breasts (1.09% vs 0.53%, p = 0.24). No significant difference was observed in positive predictive value on recall for digital mammography or screen-film mammography. Early cancer (pTis, pT1mic, pT1a) was more frequent in cancer detected with digital mammography than in that detected with screen-film mammography (41.3% vs 27.3%, p = 0.06).
Conclusion: Digital mammography may be more effective than screen-film mammography in contemporary screening practice in mobile units. The data indicate that digital mammography depicts more tumors than does screen-film mammography, especially lesions seen as microcalcifications. The potential association with improved outcome warrants further study.
Similar articles
-
Comparison of digital mammography and screen-film mammography in breast cancer screening: a review in the Irish breast screening program.AJR Am J Roentgenol. 2009 Oct;193(4):1010-8. doi: 10.2214/AJR.08.2157. AJR Am J Roentgenol. 2009. PMID: 19770323
-
Comparison of direct digital mammography, computed radiography, and film-screen in the French national breast cancer screening program.AJR Am J Roentgenol. 2014 Jan;202(1):229-36. doi: 10.2214/AJR.12.10419. AJR Am J Roentgenol. 2014. PMID: 24370149
-
Implementation of digital mammography in a population-based breast cancer screening program: effect of screening round on recall rate and cancer detection.Radiology. 2009 Jul;252(1):31-9. doi: 10.1148/radiol.2521080696. Epub 2009 May 6. Radiology. 2009. PMID: 19420316
-
Status of mammography after the Digital Mammography Imaging Screening Trial: digital versus film.Breast J. 2006 Mar-Apr;12(2):99-102. doi: 10.1111/j.1075-122X.2006.00216.x. Breast J. 2006. PMID: 16509833 Review.
-
Studies comparing screen-film mammography and full-field digital mammography in breast cancer screening: updated review.Acta Radiol. 2009 Jan;50(1):3-14. doi: 10.1080/02841850802563269. Acta Radiol. 2009. PMID: 19037825 Review.
Cited by
-
Assessing the stand-alone sensitivity of computer-aided detection with cancer cases from the Digital Mammographic Imaging Screening Trial.AJR Am J Roentgenol. 2012 Sep;199(3):W392-401. doi: 10.2214/AJR.11.7255. AJR Am J Roentgenol. 2012. PMID: 22915432 Free PMC article.
-
Preliminary clinical experience with digital breast tomosynthesis in the visualization of breast microcalcifications.J Clin Imaging Sci. 2013 Dec 31;3:65. doi: 10.4103/2156-7514.124099. eCollection 2013. J Clin Imaging Sci. 2013. PMID: 24605260 Free PMC article.
-
Interval breast cancer characteristics before, during and after the transition from screen-film to full-field digital screening mammography.BMC Cancer. 2017 May 5;17(1):315. doi: 10.1186/s12885-017-3294-5. BMC Cancer. 2017. PMID: 28476109 Free PMC article.
-
Does digital mammography in a decentralized breast cancer screening program lead to screening performance parameters comparable with film-screen mammography?Eur Radiol. 2010 Oct;20(10):2307-14. doi: 10.1007/s00330-010-1807-6. Epub 2010 May 9. Eur Radiol. 2010. PMID: 20455065
-
Impact of the Introduction of Digital Mammography in an Organized Screening Program on the Recall and Detection Rate.J Digit Imaging. 2016 Apr;29(2):235-42. doi: 10.1007/s10278-015-9843-z. J Digit Imaging. 2016. PMID: 26537932 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical