Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry
- PMID: 17898935
- DOI: 10.1007/s10792-007-9143-9
Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry
Abstract
Purpose: To compare the accuracy of central corneal thickness (CCT) measurements by the oculus pentacam scheimpflug system, with those obtained with the DGH ultrasound pachymeter (UP) and to assess the agreement between the two devices.
Methods: In a prospective study, measurement agreement was assessed in 984 eyes of 492 healthy subjects using both oculus pentacam and ultrasonic pachymetry at the Eye Consultants Center, Riyadh, Saudi Arabia.
Result: In the measurement agreement experiment, the mean CCT was 552.4 +/- 37.0 microm with oculus pentacam and 544.1 +/- 35.4 microm with ultrasonic pachymetry. Regression analysis showed a high correlation between the values obtained with both devices (r = 0.912, P < 0.001). Compared with UP, pentacam overestimated the CCT by a mean of 8.2 microm as demonstrated in a Bland-Altman plot.
Conclusion: the CCT measurements by the pentacam and UP are highly correlated. The pentacam agrees well with UP and is a reliable alternative to UP in CCT measurements.
Similar articles
-
Comparison of central corneal thickness measurements using Pentacam and ultrasonic pachymetry in post-LASIK eyes for myopia.Eur J Ophthalmol. 2010 Sep-Oct;20(5):852-7. doi: 10.1177/112067211002000507. Eur J Ophthalmol. 2010. PMID: 20815096
-
Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry.J Cataract Refract Surg. 2005 Sep;31(9):1729-35. doi: 10.1016/j.jcrs.2005.03.058. J Cataract Refract Surg. 2005. PMID: 16246776
-
[Agreement and repeatability of central corneal thickness measurement using the Pentacam and ultrasound pachymetry].Zhonghua Yan Ke Za Zhi. 2011 Feb;47(2):142-5. Zhonghua Yan Ke Za Zhi. 2011. PMID: 21426844 Chinese.
-
Comparison of corneal thickness measurements using Galilei, HR Pentacam, and ultrasound.Cornea. 2010 Oct;29(10):1091-5. doi: 10.1097/ICO.0b013e3181cf98e5. Cornea. 2010. PMID: 20628301
-
Agreement and repeatability of central thickness measurement in normal corneas using ultrasound pachymetry and the OCULUS Pentacam.Cornea. 2005 Nov;24(8):920-4. doi: 10.1097/01.ico.0000157422.01146.e9. Cornea. 2005. PMID: 16227833
Cited by
-
Repeatability and comparability of anterior segment biometry obtained by the Sirius and the Pentacam analyzers.Int Ophthalmol. 2014 Feb;34(1):27-33. doi: 10.1007/s10792-013-9780-0. Epub 2013 Apr 25. Int Ophthalmol. 2014. PMID: 23616225
-
Precision of corneal thickness measurements obtained using the scheimpflug-placido imaging and agreement with ultrasound pachymetry.J Ophthalmol. 2015;2015:328798. doi: 10.1155/2015/328798. Epub 2015 Feb 24. J Ophthalmol. 2015. PMID: 25810919 Free PMC article.
-
Measurement of anterior segment parameters in Saudi adults with myopia.Saudi J Ophthalmol. 2018 Jul-Sep;32(3):194-199. doi: 10.1016/j.sjopt.2018.04.007. Epub 2018 May 14. Saudi J Ophthalmol. 2018. PMID: 30224882 Free PMC article.
-
Normative Values for Corneal Tomography and Comparison of Both Eyes in Young Saudi Males with 20/20 Vision Using Pentacam-HR Scheimpflug Imaging.Clin Ophthalmol. 2022 Aug 13;16:2631-2637. doi: 10.2147/OPTH.S376411. eCollection 2022. Clin Ophthalmol. 2022. PMID: 35992566 Free PMC article.
-
Repeatability of intra-ocular pressure and central corneal thickness measurements provided by a non-contact method of tonometry and pachymetry.Graefes Arch Clin Exp Ophthalmol. 2011 Mar;249(3):429-34. doi: 10.1007/s00417-010-1550-3. Epub 2010 Oct 28. Graefes Arch Clin Exp Ophthalmol. 2011. PMID: 20981434
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources