Sensitivity of amplitude-integrated electroencephalography for neonatal seizure detection
- PMID: 17908764
- DOI: 10.1542/peds.2007-0514
Sensitivity of amplitude-integrated electroencephalography for neonatal seizure detection
Abstract
Background: Conventional electroencephalography remains the gold standard for the diagnosis and quantification of neonatal seizures. However, amplitude-integrated electroencephalography (aEEG) is being introduced to neonatal intensive care as an adjunct for neonatal seizure detection.
Objectives: This study's purpose was to determine the sensitivity of neonatal seizure detection in a single electroencephalogram channel (C3-->C4), used to simulate the raw signal from which aEEG is derived. We also aimed to determine the sensitivity of seizure detection by neonatologists by using aEEG and to establish those neonatal seizure characteristics that are associated with their correct detection by aEEG.
Methods: Conventional electroencephalograms with neonatal seizures were reviewed for electroencephalogram background and neonatal seizure characteristics (site of onset, duration, and peak-to-peak amplitude). The presence, duration, and peak-to-peak amplitude of each seizure were simultaneously noted in a single electroencephalogram channel (C3-->C4). aEEGs generated from this channel were reviewed for background and seizures by 6 neonatologists with varying aEEG interpretation expertise.
Results: A total of 851 neonatal seizures from 125 conventional electroencephalograms were analyzed. The patients' conceptional ages were 34 to 50 weeks. Because 94% of the conventional electroencephalograms had > or = 1 neonatal seizure visible in C3-->C4, and 78% of all neonatal seizures appeared in the C3-->C4 channel, the theoretical sensitivity of seizure detection in a single electroencephalogram channel was high. However, seizures were briefer and lower in amplitude in C3-->C4 compared with conventional electroencephalography. Neonatologists identified seizures in 22% to 57% of the 125 records of neonatal seizure. They detected 12% to 38% of the 851 individual seizures. Multivariate analysis revealed that the appearance of seizures in C3-->C4, neonatal seizure duration, seizure amplitude, seizure count per hour, and neonatologists' experience with aEEG interpretation all correlated with neonatal seizure detection.
Conclusions: Even among physicians who have extensive experience, many neonatal seizures are difficult to detect on an aEEG, especially when they are infrequent, brief, or of low amplitude.
Similar articles
-
A Comparison Of Conventional Electroencephalography With Amplitude-Integrated EEG In Detection Of Neonatal Seizures.Med Devices (Auckl). 2019 Dec 10;12:489-496. doi: 10.2147/MDER.S214662. eCollection 2019. Med Devices (Auckl). 2019. PMID: 31849541 Free PMC article. Review.
-
Inter-rater reliability of amplitude-integrated EEG for the detection of neonatal seizures.Early Hum Dev. 2020 Apr;143:105011. doi: 10.1016/j.earlhumdev.2020.105011. Epub 2020 Mar 4. Early Hum Dev. 2020. PMID: 32145502
-
Detection of subclinical electroencephalographic seizure patterns with multichannel amplitude-integrated EEG in full-term neonates.Clin Neurophysiol. 2009 Nov;120(11):1916-1922. doi: 10.1016/j.clinph.2009.08.015. Epub 2009 Sep 25. Clin Neurophysiol. 2009. PMID: 19782640
-
Accuracy of amplitude integrated EEG in a neonatal cohort.Arch Dis Child Fetal Neonatal Ed. 2010 May;95(3):F169-73. doi: 10.1136/adc.2009.165969. Arch Dis Child Fetal Neonatal Ed. 2010. PMID: 20444809
-
Amplitude-integrated electroencephalography for seizure detection in newborn infants.Semin Fetal Neonatal Med. 2018 Jun;23(3):175-182. doi: 10.1016/j.siny.2018.02.003. Epub 2018 Feb 13. Semin Fetal Neonatal Med. 2018. PMID: 29472139 Review.
Cited by
-
Continuous EEG Monitoring in Neonates: One Size Does Not Fit All.Epilepsy Curr. 2020 May 11;20(4):189-190. doi: 10.1177/1535759720923292. eCollection 2020 Jul-Aug. Epilepsy Curr. 2020. PMID: 34025224 Free PMC article. No abstract available.
-
A method for AI assisted human interpretation of neonatal EEG.Sci Rep. 2022 Jun 29;12(1):10932. doi: 10.1038/s41598-022-14894-4. Sci Rep. 2022. PMID: 35768501 Free PMC article.
-
Newborns' sleep-wake cycle development on amplitude integrated electroencephalography.World J Pediatr. 2016 Aug;12(3):327-334. doi: 10.1007/s12519-016-0026-x. Epub 2016 Jun 29. World J Pediatr. 2016. PMID: 27351568
-
A Comparison Of Conventional Electroencephalography With Amplitude-Integrated EEG In Detection Of Neonatal Seizures.Med Devices (Auckl). 2019 Dec 10;12:489-496. doi: 10.2147/MDER.S214662. eCollection 2019. Med Devices (Auckl). 2019. PMID: 31849541 Free PMC article. Review.
-
Current and Future Uses of Continuous EEG in the NICU.Front Pediatr. 2021 Nov 3;9:768670. doi: 10.3389/fped.2021.768670. eCollection 2021. Front Pediatr. 2021. PMID: 34805053 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous