Comparison of gonioscopy and anterior segment ocular coherence tomography in detecting angle closure in different quadrants of the anterior chamber angle
- PMID: 17916377
- DOI: 10.1016/j.ophtha.2007.06.030
Comparison of gonioscopy and anterior segment ocular coherence tomography in detecting angle closure in different quadrants of the anterior chamber angle
Abstract
Purpose: To compare the performance of gonioscopy and anterior segment (AS) optical coherence tomography (OCT) in detecting angle closure in the different quadrants of the anterior chamber angle (ACA).
Design: Cross-sectional observational study.
Participants: Five hundred two consecutive subjects more than 50 years of age with no previous ophthalmic problems recruited from a community clinic in Singapore.
Methods: All subjects underwent gonioscopy and AS OCT imaging in the dark. Using gonioscopy, the ACA was graded using the Scheie system by a single examiner masked to AS OCT findings.
Main outcome measures: The ACA in a particular quadrant was classified as closed if the posterior trabecular meshwork could not be seen on gonioscopy. A closed ACA on AS OCT imaging was defined by the presence of any contact between the iris and angle wall anterior to the scleral spur.
Results: After excluding eyes with poor image quality, a total of 423 right eyes were included in the analysis. A closed angle in at least 1 quadrant was observed in 59% of the eyes by AS OCT and in 33% of the eyes by gonioscopy (P<0.001), with fair agreement between the two methods (kappa = 0.40). The frequency of closed angles by AS OCT and gonioscopy were 48% versus 29% superiorly, 43% versus 22% inferiorly, 18% versus 14% nasally, and 12% versus 20% temporally, respectively. Of the 119 of 1692 quadrants that were closed on gonioscopy but open on AS OCT, a steep iris profile was present in 61 (51%) of 119 quadrants on AS OCT, and of the 276 of 1692 quadrants that were open on gonioscopy but closed on AS OCT, 196 (71%) of 276 quadrants showed short iridoangle contact on AS OCT.
Conclusions: The highest rates of closed angles on gonioscopy and AS OCT images were observed in the superior quadrant. Anterior segment OCT tended to detect more closed ACAs than gonioscopy, particularly in the superior and inferior quadrants. Variations in the iris profile and level of iridoangle contact also may explain some of the differences seen between gonioscopy and AS OCT.
Comment in
-
"Optical," not "ocular," coherence tomography.Ophthalmology. 2008 Nov;115(11):2095-6. doi: 10.1016/j.ophtha.2008.06.014. Ophthalmology. 2008. PMID: 19068377 No abstract available.
Similar articles
-
Detection of primary angle closure using anterior segment optical coherence tomography in Asian eyes.Ophthalmology. 2007 Jan;114(1):33-9. doi: 10.1016/j.ophtha.2006.05.073. Epub 2006 Oct 27. Ophthalmology. 2007. PMID: 17070597
-
Evaluation of scanning protocols for imaging the anterior chamber angle with anterior segment-optical coherence tomography.J Glaucoma. 2010 Aug;19(6):365-8. doi: 10.1097/IJG.0b013e3181bf4dd5. J Glaucoma. 2010. PMID: 19855287
-
Relationship between intraocular pressure and angle configuration: an anterior segment OCT study.Invest Ophthalmol Vis Sci. 2013 Mar 5;54(3):1650-5. doi: 10.1167/iovs.12-9986. Invest Ophthalmol Vis Sci. 2013. PMID: 23299476
-
Optical coherence tomography imaging of the anterior chamber angle.Ophthalmol Clin North Am. 2005 Sep;18(3):375-81, vi. doi: 10.1016/j.ohc.2005.05.007. Ophthalmol Clin North Am. 2005. PMID: 16054995 Review.
-
Anterior chamber angle assessment techniques.Surv Ophthalmol. 2008 May-Jun;53(3):250-73. doi: 10.1016/j.survophthal.2007.10.012. Surv Ophthalmol. 2008. PMID: 18501270 Review.
Cited by
-
Recent advances in anterior chamber angle imaging.Eye (Lond). 2020 Jan;34(1):51-59. doi: 10.1038/s41433-019-0655-0. Epub 2019 Oct 30. Eye (Lond). 2020. PMID: 31666710 Free PMC article. Review.
-
Identification of Peripheral Anterior Synechia by Corneal Deformation Using Air-Puff Dynamic Anterior Segment Optical Coherence Tomography.Front Bioeng Biotechnol. 2022 Apr 1;10:856531. doi: 10.3389/fbioe.2022.856531. eCollection 2022. Front Bioeng Biotechnol. 2022. PMID: 35433648 Free PMC article.
-
Anterior chamber angle imaging with optical coherence tomography.Eye (Lond). 2011 Mar;25(3):261-7. doi: 10.1038/eye.2010.201. Epub 2011 Jan 14. Eye (Lond). 2011. PMID: 21242985 Free PMC article.
-
Effects of diurnal, lighting, and angle-of-incidence variation on anterior segment optical coherence tomography (AS-OCT) angle metrics.BMC Ophthalmol. 2017 Mar 23;17(1):31. doi: 10.1186/s12886-017-0425-3. BMC Ophthalmol. 2017. PMID: 28335747 Free PMC article.
-
Deep Neural Network for Scleral Spur Detection in Anterior Segment OCT Images: The Chinese American Eye Study.Transl Vis Sci Technol. 2020 Mar 30;9(2):18. doi: 10.1167/tvst.9.2.18. eCollection 2020 Mar. Transl Vis Sci Technol. 2020. PMID: 32818079 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials