Post-column reactor of coaxial-gap mode for laser-induced fluorescence detection in capillary electrophoresis
- PMID: 17920070
- DOI: 10.1016/j.chroma.2007.09.046
Post-column reactor of coaxial-gap mode for laser-induced fluorescence detection in capillary electrophoresis
Abstract
A post-column reactor with coaxial-gap mode is developed for laser-induced fluorescence detection (LIF) in capillary electrophoresis (CE). The reactor can be assembled simply and conveniently, in which a thin polyimide sleeve of 10-mm length obtained from the capillary coating is used to align separation and reaction capillary with a 20 microm gap. Naphthalene-2,3-dicarboxaldehyde and 2-mercaptoethanol are used as derivatization reagents and delivered into the reaction capillary through the annulus between the separation capillary and polyimide sleeve and the gap of two capillaries by gravity. A reaction distance from the gap to detection point is 5mm. For the post-column reactor of CE-LIF, several configuration parameters are optimized, including liquid level difference between the derivatization solution and outlet buffer, annular dimension between the outer diameter of etched separation capillary and the inner diameter of polyimide sleeve, and reaction distance, etc. The detection limits in the range from 8.0x10(-8) to 1.0x10(-6) mol/L and linear calibration range more than two orders of magnitude are obtained for amino acids. The separation efficiency ranges from 1.35x10(5) to 1.67x10(5) theoretical plates.
Similar articles
-
On-line wall-free cell for laser-induced fluorescence detection in capillary electrophoresis.J Chromatogr A. 2009 May 15;1216(20):4504-9. doi: 10.1016/j.chroma.2009.03.038. Epub 2009 Mar 18. J Chromatogr A. 2009. PMID: 19329123
-
Determination of kanamycin A, amikacin and tobramycin residues in milk by capillary zone electrophoresis with post-column derivatization and laser-induced fluorescence detection.J Chromatogr B Analyt Technol Biomed Life Sci. 2009 Jan 15;877(3):333-8. doi: 10.1016/j.jchromb.2008.12.011. Epub 2008 Dec 11. J Chromatogr B Analyt Technol Biomed Life Sci. 2009. PMID: 19117808
-
On-column detection of multiphoton-excited fluorescence in CE using hyphenated cylindrical-square capillaries.Electrophoresis. 2008 Feb;29(3):734-9. doi: 10.1002/elps.200700238. Electrophoresis. 2008. PMID: 18186532
-
Post-column derivatization for fluorescence and chemiluminescence detection in capillary electrophoresis.J Pharm Biomed Anal. 1998 Sep 1;17(6-7):985-99. doi: 10.1016/s0731-7085(98)00065-x. J Pharm Biomed Anal. 1998. PMID: 9884189 Review.
-
The use of naphthalene-2,3-dicarboxaldehyde for the analysis of primary amines using high-performance liquid chromatography and capillary electrophoresis.Biomed Chromatogr. 2007 Dec;21(12):1223-39. doi: 10.1002/bmc.893. Biomed Chromatogr. 2007. PMID: 17939173 Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous