Cardiac output and cardiopulmonary responses to exercise in heart failure: application of a new bio-reactance device
- PMID: 17923354
- DOI: 10.1016/j.cardfail.2007.05.009
Cardiac output and cardiopulmonary responses to exercise in heart failure: application of a new bio-reactance device
Abstract
Background: Cardiopulmonary exercise testing (CPX) is widely used to evaluate heart failure (HF) patients, but lacks information about ventricular performance. There is a need for a noninvasive cardiac output (CO) measurement.
Methods and results: Twenty-three HF patients and 13 normals underwent CPX and CO measurements with a new, noninvasive, bioreactance technology based on assessment of relative phase shifts (d phi/d t) of electric currents injected across the thorax, heart rate, and ventricular ejection time. CO and oxygen uptake (VO2) closely paralleled each another during exercise, and peak CO was strongly correlated with peak VO2 (r = 0.73, P < .001). The relationships between peak VO2 and peak cardiac index (CI) were similar for directly measured CI (r = 0.61) and noninvasive CI (r = 0.61). The CO-VO2 relationship was also similar between the current study and previous studies from the literature, as evidenced by a similar line of best fit and 95% confidence limits. Maximal CO was significantly related to indices of ventilatory efficiency, including the VE/VCO2 slope (r = -0.47, P < .01), and the oxygen uptake efficiency slope (r = 0.67, P < .01).
Conclusion: Noninvasive measurement of CO during exercise using a novel bioreactance-based device has potentially important applications as a simple, inexpensive tool to supplement the clinical evaluation of patients with HF.
Similar articles
-
Quantification of the impaired cardiac output response to exercise in heart failure: application of a non-invasive device.J Sports Sci Med. 2009 Sep 1;8(3):344-51. eCollection 2009. J Sports Sci Med. 2009. PMID: 24149996 Free PMC article.
-
End-tidal CO2 pressure and cardiac performance during exercise in heart failure.Med Sci Sports Exerc. 2009 Jan;41(1):19-25. doi: 10.1249/MSS.0b013e318184c945. Med Sci Sports Exerc. 2009. PMID: 19092707
-
Cardiopulmonary Exercise Testing, Impedance Cardiography, and Reclassification of Risk in Patients Referred for Heart Failure Evaluation.J Card Fail. 2019 Dec;25(12):961-968. doi: 10.1016/j.cardfail.2019.08.013. Epub 2019 Aug 24. J Card Fail. 2019. PMID: 31454685
-
Evaluation of noninvasive exercise cardiac output determination in chronic heart failure patients: a proposal of a new diagnostic and prognostic method.J Cardiovasc Med (Hagerstown). 2011 Jan;12(1):19-27. doi: 10.2459/JCM.0b013e3283405c4c. J Cardiovasc Med (Hagerstown). 2011. PMID: 20975571 Review.
-
Prognostic significance and measurement of exercise-derived hemodynamic variables in patients with heart failure.J Card Fail. 2007 Oct;13(8):672-9. doi: 10.1016/j.cardfail.2007.05.004. J Card Fail. 2007. PMID: 17923361 Review.
Cited by
-
Age, sex, endurance capacity, and chronic heart failure affect central and peripheral factors of oxygen uptake measured by non-invasive and continuous technologies: support of pioneer work using invasive or non-continuous measures.Front Sports Act Living. 2023 Sep 5;5:1218948. doi: 10.3389/fspor.2023.1218948. eCollection 2023. Front Sports Act Living. 2023. PMID: 37731479 Free PMC article.
-
Quantification of the impaired cardiac output response to exercise in heart failure: application of a non-invasive device.J Sports Sci Med. 2009 Sep 1;8(3):344-51. eCollection 2009. J Sports Sci Med. 2009. PMID: 24149996 Free PMC article.
-
Rationale and design of a multi-centre randomised controlled trial of individualised cooled dialysate to prevent left ventricular systolic dysfunction in haemodialysis patients.BMC Nephrol. 2012 Jun 21;13:45. doi: 10.1186/1471-2369-13-45. BMC Nephrol. 2012. PMID: 22720738 Free PMC article. Clinical Trial.
-
Peak cardiac power measured noninvasively with a bioreactance technique is a predictor of adverse outcomes in patients with advanced heart failure.Congest Heart Fail. 2010 Nov-Dec;16(6):254-8. doi: 10.1111/j.1751-7133.2010.00187.x. Epub 2010 Oct 19. Congest Heart Fail. 2010. PMID: 21091609 Free PMC article.
-
Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique.J Clin Monit Comput. 2008 Apr;22(2):113-9. doi: 10.1007/s10877-008-9112-5. Epub 2008 Mar 14. J Clin Monit Comput. 2008. PMID: 18340540
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials
Miscellaneous