Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2008 Jan;16(1):56-63.
doi: 10.1007/s00167-007-0416-7. Epub 2007 Oct 13.

Effects of a contoured articular prosthetic device on tibiofemoral peak contact pressure: a biomechanical study

Affiliations

Effects of a contoured articular prosthetic device on tibiofemoral peak contact pressure: a biomechanical study

Christoph Becher et al. Knee Surg Sports Traumatol Arthrosc. 2008 Jan.

Abstract

Many middle-aged patients are affected by localized cartilage defects that are neither appropriate for primary, nor repeat biological repair methods, nor for conventional arthroplasty. This in vitro study aims to determine the peak contact pressure in the tibiofemoral joint with a partial femoral resurfacing device (HemiCAP, Arthrosurface Inc., Franklin, MA, USA). Peak contact pressure was determined in eight fresh-frozen cadaveric specimens using a Tekscan sensor placed in the medial compartment above the menisci. A closed loop robotic knee simulator was used to test each knee in static stance positions (5 degrees /15 degrees /30 degrees /45 degrees ) with body weight ground reaction force (GRF), 30 degrees flexion with twice the body weight (2tBW) GRF and dynamic knee-bending cycles with body weight GRF. The ground reaction force was adjusted to the living body weight of the cadaver donor and maintained throughout all cycles. Each specimen was tested under four different conditions: Untreated, flush HemiCAP implantation, 1-mm proud implantation and 20-mm defect. A paired sampled t test to compare means (significance, P < or = 0.05) was used for statistical analysis. On average, no statistically significant differences were found in any testing condition comparing the normal knee with flush device implantation. With the 1-mm proud implant, statistically significant increase of peak contact pressures of 217% (5 degrees stance), 99% (dynamic knee bending) and 90% (30 degrees stance with 2tBW) compared to the untreated condition was seen. No significant increase of peak contact pressure was evaluated with the 20-mm defect. The data suggests that resurfacing with the HemiCAP does not lead to increased peak contact pressure with flush implantation. However, elevated implantation results in increased peak contact pressure and might be biomechanically disadvantageous in an in vivo application.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Schematic drawing of the knee simulator that was used for this study
Fig. 2
Fig. 2
Different testing conditions (a Untreated; b Flush; c 1 mm; d Defect)
Fig. 3
Fig. 3
Mean peak contact pressures with one standard deviation
Fig. 4
Fig. 4
Peak contact pressure at 15° static knee stance position with single body weight of specimen No. 2. The color represents the spectrum of pressures (high pressure-red; low pressure-blue). From left to right the different testing conditions are displayed (a Untreated, b Flush, c 1 mm proud, d Defect). Marked increase of peak contact pressure at the edge of the implant to the adjacent cartilage is demonstrated (b)
Fig. 5
Fig. 5
Picture of specimen No. 2 before testing with flush implantation of the device in front view (a) and top view (b). The arrow indicates the area of peak contact pressure displayed in Fig. 4b

Similar articles

Cited by

References

    1. {'text': '', 'ref_index': 1, 'ids': [{'type': 'DOI', 'value': '10.1177/0363546505278300', 'is_inner': False, 'url': 'https://doi.org/10.1177/0363546505278300'}, {'type': 'PubMed', 'value': '16093531', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/16093531/'}]}
    2. Beck PR, Thomas AL, Farr J, Lewis PB, Cole BJ (2005) Trochlear contact pressures after anteromedialization of the tibial tubercle. Am J Sports Med 33:1710–1715 - PubMed
    1. {'text': '', 'ref_index': 1, 'ids': [{'type': 'PMC', 'value': 'PMC1888787', 'is_inner': False, 'url': 'https://pmc.ncbi.nlm.nih.gov/articles/PMC1888787/'}, {'type': 'PubMed', 'value': '16089079', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/16089079/'}]}
    2. Brand RA (2005) Joint contact stress: a reasonable surrogate for biological processes? Iowa Orthop J 25:82–94 - PMC - PubMed
    1. {'text': '', 'ref_index': 1, 'ids': [{'type': 'DOI', 'value': '10.1056/NEJM199410063311401', 'is_inner': False, 'url': 'https://doi.org/10.1056/nejm199410063311401'}, {'type': 'PubMed', 'value': '8078550', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/8078550/'}]}
    2. Brittberg M, Lindahl A, Nilsson A, Ohlsson C, Isaksson O, Peterson L (1994) Treatment of deep cartilage defects in the knee with autologous chondrocyte transplantation. N Engl J Med 331:889–895 - PubMed
    1. {'text': '', 'ref_index': 1, 'ids': [{'type': 'DOI', 'value': '10.1002/jor.1100090412', 'is_inner': False, 'url': 'https://doi.org/10.1002/jor.1100090412'}, {'type': 'PubMed', 'value': '2045983', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/2045983/'}]}
    2. Brown TD, Pope DF, Hale JE, Buckwalter JA, Brand RA (1991) Effects of osteochondral defect size on cartilage contact stress. J Orthop Res 9:559–567 - PubMed
    1. {'text': '', 'ref_index': 1, 'ids': [{'type': 'DOI', 'value': '10.1097/00003086-197201000-00033', 'is_inner': False, 'url': 'https://doi.org/10.1097/00003086-197201000-00033'}, {'type': 'PubMed', 'value': '5011034', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/5011034/'}]}
    2. Convery FR, Akeson WH, Keown GH (1972) The repair of large osteochondral defects. An experimental study in horses. Clin Orthop Relat Res 82:253–262 - PubMed