Pulse oximetry: basic principles and applications in aerospace medicine
- PMID: 17955947
- DOI: 10.3357/asem.2087.2007
Pulse oximetry: basic principles and applications in aerospace medicine
Abstract
Introduction: Pulse oximeters are reliable, objective, and noninvasive monitors that have broad application in aerospace medicine. New technology enables pulse oximeters to perform well in adverse environments and measure additional parameters. Small, battery-powered devices can be used to monitor oxyhemoglobin saturation while in flight.
Theory of operation: Pulse oximeters use spectrophotometry to measure the ratio of oxyhemoglobin (Hbo2) to reduced hemoglobin (Hb) in arterial blood. This value is displayed as oxyhemoglobin saturation (Spo2). A plethysmographic waveform that resembles arterial waveform is also frequently displayed and may indicate relative changes in perfusion and blood volume. Loss and subsequent reappearance of this waveform during occlusion with a cuff has been used to measure systolic blood pressure during helicopter flight.
Applications: Accurate determination of oxygen saturation requires a high quality arterial signal and is limited by errors resulting from calibration, motion and vibration, and dyshemoglobinemias. Vasoconstriction may result in decreased pulse amplitude and also impair accurate measurement. Conventional fingertip probes may interfere with the performance of required duties, while helmets and other restrictive clothing can impede the use of sensors on the forehead or ear. Recently introduced devices answer some of these limitations and enable measurement of additional parameters. For example, new probe designs permit more freedom of movement and include a contactless camera and a sensor that fits around a finger like an ordinary ring. This article explains the theory of operation and limitations of pulse oximetry, offers an update on new technology, and discusses applications of this technology in aerospace medicine.
Comment in
-
Pulse oximetry: basic principles and applications in aerospace medicine.Aviat Space Environ Med. 2008 Apr;79(4):444; author reply 444. doi: 10.3357/asem.86081.2008. Aviat Space Environ Med. 2008. PMID: 18457305 No abstract available.
Similar articles
-
A review of the principles of pulse oximetry and accuracy of pulse oximeter estimates during exercise.Phys Ther. 1994 Jan;74(1):40-9. doi: 10.1093/ptj/74.1.40. Phys Ther. 1994. PMID: 8265727 Review.
-
Influence of digital clubbing on oxygen saturation measurements by pulse-oximetry in cystic fibrosis patients.J Cyst Fibros. 2006 May;5(2):125-8. doi: 10.1016/j.jcf.2006.01.007. Epub 2006 Feb 28. J Cyst Fibros. 2006. PMID: 16504602
-
The effects of motion artifact and low perfusion on the performance of a new generation of pulse oximeters in volunteers undergoing hypoxemia.Respir Care. 2002 Jan;47(1):48-60. Respir Care. 2002. PMID: 11749687
-
Next-generation pulse oximetry. Focusing on Masimo's signal extraction technology.Health Devices. 2000 Oct;29(10):349-70. Health Devices. 2000. PMID: 11094776
-
Pulse oximetry: a review of the theory, accuracy, and clinical applications.Obstet Gynecol. 1989 Sep;74(3 Pt 2):541-6. Obstet Gynecol. 1989. PMID: 2668828 Review.
Cited by
-
Persistent low oxygen saturation on polysomnogram.J Clin Sleep Med. 2013 Sep 15;9(9):967-9. doi: 10.5664/jcsm.3006. J Clin Sleep Med. 2013. PMID: 23997710 Free PMC article. No abstract available.
-
In-flight medical emergencies: time for a registry?Crit Care. 2009;13(1):121. doi: 10.1186/cc7715. Epub 2009 Feb 25. Crit Care. 2009. PMID: 19291266 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials