Stability-indicating reversed-phase liquid chromatographic method for simultaneous determination of simvastatin and ezetimibe from their combination drug products
- PMID: 17955968
Stability-indicating reversed-phase liquid chromatographic method for simultaneous determination of simvastatin and ezetimibe from their combination drug products
Abstract
A simple, precise, and rapid stability-indicating reversed-phase column liquid chromatographic (RP-LC) method has been developed and subsequently validated for simultaneous estimation of simvastatin (SIM) and ezetimibe (EZE) from their combination drug product. The proposed RP-LC method utilizes a LiChrospher 100 C18, 5 microm, 250 x 4.0 mm id column at ambient temperature; optimum mobile phase consisting of acetonitrile-water-methanol (60 + 25 + 15, v/v/v) with apparent pH adjusted to 4.0 +/- 0.1; mobile phase flow rate of 1.5 mL/min; and ultraviolet detection at 238 nm. SIM, EZE, and their combination drug product were exposed to thermal, photolytic, hydrolytic, and oxidative stress conditions, and the stressed samples were analyzed by the proposed method. There were no other coeluting, interfering peaks from excipients, impurities, or degradation products due to variable stress conditions, and the method is specific for the estimation of SIM and EZE in the presence of degradation products. The described method was linear over the range of 1-80 and 3-80 microg/mL for SIM and EZE, respectively. The mean recoveries were 99.17 and 100.43% for SIM and EZE, respectively. The intermediate precision data were obtained under different experimental conditions, and the calculated value of the coefficient of variation was found to be less than the critical value. The proposed method can be useful in the quality control of bulk manufacturing and pharmaceutical dosage forms.
Similar articles
-
Stability-indicating reversed-phase liquid chromatographic method for simultaneous determination of atorvastatin and ezetimibe from their combination drug products.J AOAC Int. 2007 Nov-Dec;90(6):1539-46. J AOAC Int. 2007. PMID: 18193730
-
Rapid and sensitive simultaneous determination of ezetimibe and simvastatin from their combination drug products by monolithic silica high-performance liquid chromatographic column.J Pharm Biomed Anal. 2009 Oct 15;50(3):527-34. doi: 10.1016/j.jpba.2009.05.002. Epub 2009 May 13. J Pharm Biomed Anal. 2009. PMID: 19487095
-
Simultaneous liquid chromatographic determination of ezetimibe and simvastatin in pharmaceutical products.J AOAC Int. 2007 Nov-Dec;90(6):1566-72. J AOAC Int. 2007. PMID: 18193733
-
Ezetimibe: A Review of Analytical Methods for the Drug Substance, Pharmaceutical Formulations and Biological Matrices.Crit Rev Anal Chem. 2022;52(5):1078-1093. doi: 10.1080/10408347.2020.1857222. Epub 2020 Dec 21. Crit Rev Anal Chem. 2022. PMID: 33347374 Review.
-
Development of validated stability-indicating assay methods--critical review.J Pharm Biomed Anal. 2002 Jun 15;28(6):1011-40. doi: 10.1016/s0731-7085(02)00047-x. J Pharm Biomed Anal. 2002. PMID: 12049968 Review.
Cited by
-
Stability-indicating RP-HPLC Method for the Simultaneous Determination of Sitagliptin and Simvastatin in Tablets.Indian J Pharm Sci. 2014 Sep;76(5):407-14. Indian J Pharm Sci. 2014. PMID: 25425754 Free PMC article.