The accuracy of the Biox 3700 pulse oximeter in patients receiving vasoactive therapy
- PMID: 1797894
- DOI: 10.1007/BF01690773
The accuracy of the Biox 3700 pulse oximeter in patients receiving vasoactive therapy
Abstract
The accuracy of pulse oximetry for the prediction of oxygen saturation of arterial blood in patients with shock has been hardly studied. This study was undertaken to determine if O2 arterial saturation estimated by the Biox 3700 pulse oximeter (SpO2) with an ear probe could reliably substitute for the measurements of O2 saturation (HbO2) with an IL-282 Co-Oximeter in samples of arterial blood obtained from 24 caucasian patients. All patients were mechanically ventilated and needed vasoactive drugs (dopamine and/or dobutamine). Of 24 patients 13 had shock: cardiogenic (n = 6) and septic (n = 7). The mean difference between HbO2 and SpO2 was 2.49% +/- 4.24, with a 95% confidence interval of 0.7% to 4.3% (p = 0.009). There were clinically important differences between both methods since in 9 of 24 patients (37%) SpO2 values were at least 4% lower or greater than HbO2. This disagreement was also apparent in patients with (6/13, 46%) or without shock (3/11, 27%). To conclude, pulse oximetry is not always a sufficiently reliable method to predict HbO2 in patients with or without shock treated with vasoactive drugs.
Similar articles
-
Pulse oximetry in newborn infants with birth weights of 620 to 4285 grams receiving dopamine and dobutamine.J Perinatol. 1996 Jan-Feb;16(1):31-4. J Perinatol. 1996. PMID: 8869537
-
An evaluation of pulse oximetry in prehospital care.Ann Emerg Med. 1991 Aug;20(8):887-91. doi: 10.1016/s0196-0644(05)81432-5. Ann Emerg Med. 1991. PMID: 1854074
-
The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study.BMC Emerg Med. 2010 May 5;10:9. doi: 10.1186/1471-227X-10-9. BMC Emerg Med. 2010. PMID: 20444248 Free PMC article.
-
A review of the principles of pulse oximetry and accuracy of pulse oximeter estimates during exercise.Phys Ther. 1994 Jan;74(1):40-9. doi: 10.1093/ptj/74.1.40. Phys Ther. 1994. PMID: 8265727 Review.
-
Advanced Uses of Pulse Oximetry for Monitoring Mechanically Ventilated Patients.Anesth Analg. 2017 Jan;124(1):62-71. doi: 10.1213/ANE.0000000000001283. Anesth Analg. 2017. PMID: 27183375 Review.
Cited by
-
Pulse oximetry and high-dose vasopressors: a comparison between forehead reflectance and finger transmission sensors.Intensive Care Med. 2012 Oct;38(10):1718-22. doi: 10.1007/s00134-012-2659-0. Epub 2012 Aug 7. Intensive Care Med. 2012. PMID: 22868275
-
Impact of COVID-19 on the association between pulse oximetry and arterial oxygenation in patients with acute respiratory distress syndrome.Sci Rep. 2022 Jan 27;12(1):1462. doi: 10.1038/s41598-021-02634-z. Sci Rep. 2022. PMID: 35087122 Free PMC article.
-
Accuracy of pulse oximetry in detection of oxygen saturation in patients admitted to the intensive care unit of heart surgery: comparison of finger, toe, forehead and earlobe probes.BMC Nurs. 2018 Apr 17;17:15. doi: 10.1186/s12912-018-0283-1. eCollection 2018. BMC Nurs. 2018. PMID: 29692684 Free PMC article.
-
Donor preoperative oxygen delivery and post-extubation hypoxia impact donation after circulatory death hypoxic cholangiopathy.World J Gastroenterol. 2016 Mar 28;22(12):3392-403. doi: 10.3748/wjg.v22.i12.3392. World J Gastroenterol. 2016. PMID: 27022221 Free PMC article.
-
Assessing analgesia equivalence and appetite following alfaxalone- or ketamine-based injectable anesthesia for feline castration as an example of enhanced recovery after surgery.J Feline Med Surg. 2018 Feb;20(2):73-82. doi: 10.1177/1098612X17693517. Epub 2017 Feb 1. J Feline Med Surg. 2018. PMID: 28682644 Free PMC article.