Effect of positive expiratory pressure and type of tracheal cuff on the incidence of aspiration in mechanically ventilated patients in an intensive care unit
- PMID: 18007268
- DOI: 10.1097/01.CCM.0000297888.82492.31
Effect of positive expiratory pressure and type of tracheal cuff on the incidence of aspiration in mechanically ventilated patients in an intensive care unit
Abstract
Objective: To test the effects of positive expiratory pressure on the leakage of fluid around cuffs of different tracheal tubes, in mechanically ventilated patients and in a benchtop model.
Design: Randomized clinical trial and experimental in vitro study.
Setting: Intensive care unit of a university hospital.
Patients: Forty patients recovering in the intensive care unit were ventilated in volume-controlled mode. Twenty patients were randomly intubated with Hi-Lo tubes (HL group), whereas the remaining 20 subjects were intubated with SealGuard tubes (SG group).
Interventions: Immediately after intubation and cuff inflation with 30 cm H2O, Evans blue was applied onto the cephalic surface of the tracheal tube cuff. A 5-cm H2O positive expiratory pressure was used during the first 5 hrs of stay, and thereafter it was removed. Bronchoscopy verified whether the dye leaked around the cuff. The experiment lasted 12 hrs. Leakage was also tested in vitro with the same tracheal tubes with incremental level of positive expiratory pressure.
Measurements and main results: At 1 hr, 5 hrs, and thereafter hourly until 12 hrs, bronchoscopy was used to test the presence of dye on the trachea caudal to the cuff. At the fifth hour, two patients of the HL group failed the test. One hour after positive expiratory pressure removal, all subjects in group HL exhibited a dyed lower trachea. On the other hand, one patient in group SG presented a leak at the eighth hour, and at the 12th hour three of them were still sealed. In vitro, the same level of positive expiratory pressure delayed the passage of dye around the cuff; after 30 mins positive expiratory pressure was removed, and in 10 mins all dye leaked only in the Hi-Lo tube.
Conclusions: We found that 5 cm H2O positive expiratory pressure was effective in delaying the passage of fluid around the cuffs of tracheal tubes both in vivo and in vitro. The SealGuard tube proved to be more resistant to leakage than Hi-Lo.
Comment in
-
Subglottic secretions and Sutton's law: a simple and effective approach.Crit Care Med. 2008 Feb;36(2):623-4. doi: 10.1097/01.CCM.0000299846.73263.37. Crit Care Med. 2008. PMID: 18216615 No abstract available.
-
The effect of positive expiratory pressure and tracheal tube cuff type on pulmonary aspiration.Crit Care Med. 2008 May;36(5):1691-2; author reply 1692. doi: 10.1097/CCM.0b013e3181710296. Crit Care Med. 2008. PMID: 18448951 No abstract available.
-
Does polyurethane impact endotracheal cuff pressure?Crit Care Med. 2008 Jul;36(7):2219; author reply 2219-20. doi: 10.1097/CCM.0b013e31817c47d9. Crit Care Med. 2008. PMID: 18594246 No abstract available.
Similar articles
-
Automatic control of tracheal tube cuff pressure in ventilated patients in semirecumbent position: a randomized trial.Crit Care Med. 2007 Jun;35(6):1543-9. doi: 10.1097/01.CCM.0000266686.95843.7D. Crit Care Med. 2007. PMID: 17452937 Clinical Trial.
-
Prevention of silent aspiration due to leaks around cuffs of endotracheal tubes.Anesth Analg. 1986 Jul;65(7):777-80. Anesth Analg. 1986. PMID: 3717617 Clinical Trial.
-
Performance of the PneuX System: A Bench Study Comparison With 4 Other Endotracheal Tube Cuffs.Respir Care. 2017 Jan;62(1):102-112. doi: 10.4187/respcare.04996. Epub 2016 Nov 22. Respir Care. 2017. PMID: 27879379
-
The LoTrach system: its role in the prevention of ventilator-associated pneumonia.Nurs Crit Care. 2008 Sep-Oct;13(5):260-8. doi: 10.1111/j.1478-5153.2008.00286.x. Nurs Crit Care. 2008. PMID: 18816312 Review.
-
Microaspiration in intubated critically ill patients: diagnosis and prevention.Infect Disord Drug Targets. 2011 Aug;11(4):413-23. doi: 10.2174/187152611796504827. Infect Disord Drug Targets. 2011. PMID: 21679139 Review.
Cited by
-
The effect of tracheal tube size on air leak around the cuffs.Korean J Anesthesiol. 2011 Jul;61(1):24-9. doi: 10.4097/kjae.2011.61.1.24. Epub 2011 Jul 21. Korean J Anesthesiol. 2011. PMID: 21860747 Free PMC article.
-
Insights on the role of antimicrobial cuffed endotracheal tubes in preventing transtracheal transmission of VAP pathogens from an in vitro model of microaspiration and microbial proliferation.Biomed Res Int. 2014;2014:120468. doi: 10.1155/2014/120468. Epub 2014 Apr 10. Biomed Res Int. 2014. PMID: 24818125 Free PMC article.
-
Technologic advances in endotracheal tubes for prevention of ventilator-associated pneumonia.Chest. 2012 Jul;142(1):231-238. doi: 10.1378/chest.11-2420. Chest. 2012. PMID: 22796845 Free PMC article. Review.
-
Tracheal fluid leakage in benchtop trials: comparison of static versus dynamic ventilation model with and without lubrication.J Anesth. 2010 Apr;24(2):247-52. doi: 10.1007/s00540-010-0871-z. Epub 2010 Feb 23. J Anesth. 2010. PMID: 20174988
-
Design considerations for development of cuffed endotracheal tube for small airways.Paediatr Anaesth. 2025 Feb;35(2):91-98. doi: 10.1111/pan.15033. Epub 2024 Nov 19. Paediatr Anaesth. 2025. PMID: 39560156 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical