Initial clinical experience with a new end graft holder for anastomosis in coronary surgery
- PMID: 18018605
- DOI: 10.1007/s11748-007-0155-1
Initial clinical experience with a new end graft holder for anastomosis in coronary surgery
Abstract
Objective: We assessed the feasibility and effectiveness of a novel end graft holder for coronary artery bypass grafting (CABG) and evaluated anastomotic patency and early clinical results.
Methods: The end graft holder was applied to 45 consecutive patients. Operative characteristics were off-pump CABG in 22.2%, emergency in 28.9%, and concomitant cardiac surgery in 13.3%.
Results: The device was used safely without graft injury or inadequate gripping on grafts. Postoperative angiography showed that the patency rate of distal anastomosis was 96.7% (arterial, 100%; venous, 94%). All proximal aortic and composite graft anastomoses were patent without stenosis. The rate of 30-day major adverse cardiac and cerebrovascular events was 13.3% (operative deaths, 3; repeated CABG, 1; percutaneous coronary arterial intervention, 1; and cerebral infarction, 1). None of the elective patients died during hospitalization.
Conclusion: Our initial clinical experience demonstrated that the new end graft holder was safe, reliable, and effective during CABG. The excellent fixation and visualization of the graft with the device might be particularly beneficial for off-pump CABG or for teaching trainees.
Similar articles
-
Graft fixation with a side graft holder for sequential and composite graft anastomosis in coronary artery bypass surgery.Interact Cardiovasc Thorac Surg. 2009 Dec;9(6):939-42. doi: 10.1510/icvts.2009.215608. Epub 2009 Sep 25. Interact Cardiovasc Thorac Surg. 2009. PMID: 19783547
-
Initial experience of an automated anastomotic distal device in off-pump CABG.Heart Surg Forum. 2004;7(5):E360-3. doi: 10.1532/HSF98.20041060. Heart Surg Forum. 2004. PMID: 15799903 Clinical Trial.
-
Application of a new suture material called the U-Clip for composite and sequential grafting with off-pump coronary bypass surgery.Heart Surg Forum. 2006;9(6):E861-5. doi: 10.1532/HSF98.20061084. Heart Surg Forum. 2006. PMID: 17060041 Clinical Trial.
-
Does off-pump coronary artery bypass graft surgery have a beneficial effect on long-term mortality and morbidity compared with on-pump coronary artery bypass graft surgery?Interact Cardiovasc Thorac Surg. 2014 Jul;19(1):149-59. doi: 10.1093/icvts/ivu075. Epub 2014 Mar 21. Interact Cardiovasc Thorac Surg. 2014. PMID: 24659552 Review.
-
Anastomotic devices for coronary artery bypass grafting: technological options and potential pitfalls.Comput Biol Med. 2007 Oct;37(10):1384-93. doi: 10.1016/j.compbiomed.2006.11.005. Epub 2007 Jan 19. Comput Biol Med. 2007. PMID: 17240365 Review.
References
Publication types
MeSH terms
LinkOut - more resources