Comparison of BD GeneOhm methicillin-resistant Staphylococcus aureus (MRSA) PCR versus the CHROMagar MRSA assay for screening patients for the presence of MRSA strains
- PMID: 18032616
- PMCID: PMC2224301
- DOI: 10.1128/JCM.02130-07
Comparison of BD GeneOhm methicillin-resistant Staphylococcus aureus (MRSA) PCR versus the CHROMagar MRSA assay for screening patients for the presence of MRSA strains
Abstract
We compared the BD GeneOhm methicillin-resistant Staphylococcus aureus (MRSA) real-time PCR assay with the CHROMagar MRSA assay for the detection of MRSA in 286 nasal surveillance specimens. Compared with the CHROMagar MRSA assay, PCR had sensitivity, specificity, positive predictive value, and negative predictive values of 100%, 98.6%, 95.8%, and 100%, respectively. The mean PCR turnaround time was 14.5 h.
References
-
- Bishop, E. J., E. A. Grabsch, S. A. Ballard, B. Mayall, S. Xie, R. Martin, and M. L. Grayson. 2006. Concurrent analysis of nose and groin swab specimens by the IDI-MRSA PCR assay is comparable to analysis by individual-specimen PCR and routine culture assays for detection of colonization by methicillin-resistant Staphylococcus aureus. J. Clin. Microbiol. 442904-2908. - PMC - PubMed
-
- Flayhart, D., J. F. Hindler, D. A. Bruckner, G. Hall, R. K. Shrestha, S. A. Vogel, S. S. Richter, W. Howard, R. Walther, and K. C. Carroll. 2005. Multicenter evaluation of BBL CHROMagar MRSA medium for direct detection of methicillin-resistant Staphylococcus aureus from surveillance cultures of the anterior nares. J. Clin. Microbiol. 435536-5540. - PMC - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical