MOSFET detectors in quality assurance of tomotherapy treatments
- PMID: 18037519
- DOI: 10.1016/j.radonc.2007.10.025
MOSFET detectors in quality assurance of tomotherapy treatments
Abstract
Background and purpose: The purpose of this work was to characterize metal oxide semiconductor field-effect transistors (MOSFETs) in a 6 MV conventional linac and investigate their use for quality assurance of radiotherapy treatments with a tomotherapy Hi-Art unit.
Materials and methods: High sensitivity and standard sensitivity MOSFETs were first calibrated and then tested for reproducibility, field size dependence, and accuracy of measuring surface dose in a 6 MV beam as well as in a tomotherapy Hi-Art unit. In vivo measurements were performed on both a RANDO phantom and several head and neck cancer patients treated with tomotherapy and compared to TLD measurements and treatment plan doses to evaluate the performance of MOSFETs in a high gradient radiation field.
Results: The average calibration factor found was 0.345+/-2.5%cGy/mV for the high sensitivity MOSFETs tested and 0.901+/-2.4%cGy/mV for the standard sensitivity MOSFETs. MOSFET measured surface doses had an average agreement with ion chamber measurements of 1.55% for the high sensitivity MOSFET and 5.23% for the standard sensitivity MOSFET when averaged over all trials and field sizes tested. No significant dependence on field size was found for the standard sensitivity MOSFETs, however a maximum difference of 5.34% was found for the high sensitivity MOSFET calibration factors in the field sizes tested. Measurements made with MOSFETS on head and neck patients treated on a tomotherapy Hi-Art unit had an average agreement of (3.26+/-0.03)% with TLD measurements, however the average of the absolute difference between the MOSFET measurements and the treatment plan skin doses was (12.2+/-7.5)%. The MOSFET measured patient skin doses also had good reproducibility, with inter-fraction deviations ranging from 1.4% to 6.6%. Similar results were found from trials using a RANDO phantom.
Conclusions: The MOSFETs performed well when used in the tomotherapy Hi-Art unit and did not increase the overall treatment set-up time when used for patient measurements. It was found that MOSFETs are suitable detectors for surface dose measurements in both conventional beam and tomotherapy treatments and they can provide valuable skin dose information in areas where the treatment planning system may not be accurate.
Similar articles
-
Characterization of metal oxide field-effect transistors for first helical tomotherapy Hi-Art II unit in India.J Cancer Res Ther. 2009 Oct-Dec;5(4):284-9. doi: 10.4103/0973-1482.59911. J Cancer Res Ther. 2009. PMID: 20160363
-
Skin dose measurements using MOSFET and TLD for head and neck patients treated with tomotherapy.Appl Radiat Isot. 2009 Sep;67(9):1683-5. doi: 10.1016/j.apradiso.2009.03.008. Epub 2009 Mar 24. Appl Radiat Isot. 2009. PMID: 19369084
-
Clinical implementation of MOSFET detectors for dosimetry in electron beams.Radiother Oncol. 2006 Sep;80(3):288-95. doi: 10.1016/j.radonc.2006.07.002. Epub 2006 Aug 17. Radiother Oncol. 2006. PMID: 16916563
-
[Verification of the dose delivered to the patient by means of TLD, SC, PID. What future?].Cancer Radiother. 2003 Nov;7 Suppl 1:78s-84s. Cancer Radiother. 2003. PMID: 15124548 Review. French.
-
Nanometer size field effect transistors for terahertz detectors.Nanotechnology. 2013 May 31;24(21):214002. doi: 10.1088/0957-4484/24/21/214002. Epub 2013 Apr 25. Nanotechnology. 2013. PMID: 23618776 Review.
Cited by
-
Peripheral dose measurement in high-energy photon radiotherapy with the implementation of MOSFET.World J Radiol. 2010 Nov 28;2(11):434-9. doi: 10.4329/wjr.v2.i11.434. World J Radiol. 2010. PMID: 21179311 Free PMC article.
-
Evaluation of the usefulness of a MOSFET detector in an anthropomorphic phantom for 6-MV photon beam.Radiol Phys Technol. 2010 Jul;3(2):104-12. doi: 10.1007/s12194-010-0084-z. Epub 2010 Feb 4. Radiol Phys Technol. 2010. PMID: 20821083
-
Acute Toxicity From Breast Cancer Radiation Using Helical Tomotherapy With a Simultaneous Integrated Boost.Technol Cancer Res Treat. 2016 Apr;15(2):257-65. doi: 10.1177/1533034615574387. Epub 2015 Mar 16. Technol Cancer Res Treat. 2016. PMID: 25780060 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical